Laboratory Evaluation of the Influence of Decentration and Pupil Size on the Optical Performance of a Monofocal, Bifocal, and Trifocal Intraocular Lens
- PMID: 29227508
- DOI: 10.3928/1081597X-20171004-02
Laboratory Evaluation of the Influence of Decentration and Pupil Size on the Optical Performance of a Monofocal, Bifocal, and Trifocal Intraocular Lens
Abstract
Purpose: To assess the influence of decentration and aperture size on the optical quality of different intraocular lenses (IOLs) of the same material, body design, and refractive power using standardized optical bench testing.
Methods: Using an optical bench set-up, an aspheric monofocal (CT ASPHINA 409M; Carl Zeiss Meditec, Jena, Germany), an aspheric diffractive bifocal (AT LISA 809M; Carl Zeiss Meditec), and an aspheric diffractive trifocal (AT LISA 839M; Carl Zeiss Meditec) intraocular lens (IOL) were evaluated, each with the same distance power, body design, and material. Modulation transfer function (MTF) values were measured at spatial frequencies of 50 lp/mm and aperture sizes of 3 and 4.5 mm. Each IOL was measured while centered, then decentered by 0.25, 0.5, 0.75, and 1 mm.
Results: MTF values for the monofocal IOL at far focus with 3- and 4.5-mm aperture size were 0.80/0.80 with maximum reduction to 0.77/0.73 for 1-mm decentration, respectively. Centered IOL MTFs of the bifocal and trifocal IOLs were lower for the far focus at 0.46/0.41 and 0.39/0.26, with reduction at 1-mm decentration to 0.35/0.25 and 0.25/0.18, respectively. Values for near focus of the bifocal and trifocal IOLs reduced from 0.27/0.31 and 0.19/0.18 to 0.2/0.21 and 0.12/0.13, respectively. The trifocal intermediate focus MTF reduced from 0.15/0.10 to 0.12/0.08. MTF values of all three lenses decreased significantly under all conditions with decentration of 0.5 to 0.75 mm.
Conclusions: Monofocal lenses were least negatively affected by decentration, with mean optical quality reduction of less than 10% for 1-mm decentration at physiological pupil sizes. For diffractive bifocal and trifocal lenses, optical quality at all distances was significantly reduced if decentration exceeded 0.75 mm, with intermediate focus showing the least reduction. [J Refract Surg. 2017;33(12):808-812.].
Copyright 2017, SLACK Incorporated.
Similar articles
-
In vitro optical quality measurements of three intraocular lens models having identical platform.BMC Ophthalmol. 2017 Jun 29;17(1):108. doi: 10.1186/s12886-017-0460-0. BMC Ophthalmol. 2017. PMID: 28662629 Free PMC article.
-
Effect of Decentration on the Optical Quality of Two Intraocular Lenses.Optom Vis Sci. 2016 Dec;93(12):1552-1559. doi: 10.1097/OPX.0000000000001004. Optom Vis Sci. 2016. PMID: 27776082
-
Effect of decentration and tilt on the in vitro optical quality of monofocal and trifocal intraocular lenses.Graefes Arch Clin Exp Ophthalmol. 2024 Oct;262(10):3229-3242. doi: 10.1007/s00417-024-06490-1. Epub 2024 Apr 21. Graefes Arch Clin Exp Ophthalmol. 2024. PMID: 38643424
-
[Multifocal intraocular lenses--an assessment of current status].Klin Monbl Augenheilkd. 1993 Jul;203(1):19-33. doi: 10.1055/s-2008-1045645. Klin Monbl Augenheilkd. 1993. PMID: 8411889 Review. German.
-
AcrySof IQ PanOptix Intraocular Lens Versus Extended Depth of Focus Intraocular Lens and Trifocal Intraocular Lens: A Clinical Overview.Asia Pac J Ophthalmol (Phila). 2019 Jul-Aug;8(4):335-349. doi: 10.1097/APO.0000000000000253. Asia Pac J Ophthalmol (Phila). 2019. PMID: 31403494 Free PMC article. Review.
Cited by
-
Tilt and decentration with various intraocular lenses: A narrative review.World J Clin Cases. 2022 Apr 26;10(12):3639-3646. doi: 10.12998/wjcc.v10.i12.3639. World J Clin Cases. 2022. PMID: 35647149 Free PMC article. Review.
-
Influence of decentration of plate-haptic toric intraocular lens on postoperative visual quality.BMC Ophthalmol. 2023 Jul 20;23(1):332. doi: 10.1186/s12886-023-03061-6. BMC Ophthalmol. 2023. PMID: 37474888 Free PMC article.
-
Refractive Outcomes after Cataract Surgery.Diagnostics (Basel). 2022 Jan 19;12(2):243. doi: 10.3390/diagnostics12020243. Diagnostics (Basel). 2022. PMID: 35204334 Free PMC article. Review.
-
[Unilateral implantation of a new extended range of vision intraocular lens in a young patient].Ophthalmologe. 2022 Jan;119(1):65-70. doi: 10.1007/s00347-020-01285-1. Epub 2020 Dec 10. Ophthalmologe. 2022. PMID: 33301065 Free PMC article. German. No abstract available.
-
A Comparative Analysis of the Effects of Misaligning Different Trifocal Intraocular Lenses.J Clin Med. 2024 Dec 31;14(1):187. doi: 10.3390/jcm14010187. J Clin Med. 2024. PMID: 39797271 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous