Gross Trunnion Failure of a Cobalt-Chromium Femoral Head on a Titanium Stem at Midterm Follow-up: A Report of 3 Cases
- PMID: 29252750
- DOI: 10.2106/JBJS.CC.16.00054
Gross Trunnion Failure of a Cobalt-Chromium Femoral Head on a Titanium Stem at Midterm Follow-up: A Report of 3 Cases
Abstract
Case: Three patients underwent uncomplicated primary total hip arthroplasty with cobalt-chromium femoral heads (36+5 mm) on titanium V40 tapers. At 6 to 9 years of follow-up, severe effects of corrosion at the trunnion were noted in all 3 patients, along with elevated levels of serum cobalt ions and normal levels of serum chromium ions. Gross trunnion failure, apparently caused by corrosion, required femoral stem revision in all of the patients.
Conclusion: Decreased neck diameter, longer trunnion length, and large-sized cobalt-chromium heads are possible contributors to early failure after primary total hip arthroplasty due to trunnionosis. Surgeons should be mindful of trunnionosis as a cause of pain and a mechanism of failure following total hip arthroplasty, and serum metal ions should be monitored in these patients.
References
-
- Archibeck MJ, Cummins T, Carothers J, Junick DW, White RE Jr. A comparison of two implant systems in restoration of hip geometry in arthroplasty. Clin Orthop Relat Res. 2011 Feb;469(2):443-6.
-
- Pastides PS, Dodd M, Sarraf KM, Willis-Owen CA. Trunnionosis: a pain in the neck. World J Orthop. 2013;4(4):161-6. Epub 2013 Oct 18.
-
- Gilbert JL, Buckley CA, Jacobs JJ. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling. J Biomed Mater Res. 1993 Dec;27(12):1533-44.
-
- Lohmann CH, Singh G, Willert HG, Buchhorn GH. Metallic debris from metal-on-metal total hip arthroplasty regulates periprosthetic tissues. World J Orthop. 2014 Nov 18;5(5):660-6. Epub 2014 Nov 18.
-
- Lombardi AV Jr, Barrack RL, Berend KR, Cuckler JM, Jacobs JJ, Mont MA, Schmalzried TP. The Hip Society: algorithmic approach to diagnosis and management of metal-on-metal arthroplasty. J Bone Joint Surg Br. 2012 Nov;94(11)(Suppl A):14-8.