Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Multicenter Study
. 2018 Feb-Mar:99-100:1-4.
doi: 10.1016/j.jcv.2017.12.006. Epub 2017 Dec 13.

Performance evaluation of the Aptima HSV-1 and 2 assay for the detection of HSV in cutaneous and mucocutaneous lesion specimens

Affiliations
Multicenter Study

Performance evaluation of the Aptima HSV-1 and 2 assay for the detection of HSV in cutaneous and mucocutaneous lesion specimens

Soya S Sam et al. J Clin Virol. 2018 Feb-Mar.

Abstract

Background: Timely and precise laboratory diagnosis of Herpes simplex viruses (HSV) is required to guide clinical management.

Objectives: The study evaluated limit of detection (LOD) and performance characteristics of the Aptima HSV 1 & 2 assay in comparison to four assays.

Study design: The multi-center study compared qualitative detection of HSV-1 and 2 by the Aptima HSV-1 and 2 assay (Hologic) to ELVIS culture, Lyra Direct (Quidel), AmpliVue (Quidel) and a laboratory developed test (LDT). LOD was performed using VTM and STM diluted viral concentrations and clinical performance was evaluated using 505 swab specimens.

Results: The Aptima LOD studies performed showed a lower detection limit for STM specimens as 1450 copies/mL and 430 copies/mL for HSV1 and HSV-2 respectively; the LOD for VTM specimens was 9370 copies/mL and 8045 copies/mL for HSV-1 and HSV-2 respectively. When the assays were analyzed based on the positive consensus result established the Aptima had 95% of percent positive agreement (PPA) and 100% negative percent agreement (NPA) for the HSV-1. For the HSV-2, the PPA and NPA for Aptima were 96% and 100% respectively. AmpliVue had 1.8% invalid rate, while Lyra had no invalid results but an inhibition rate of 0.8%. Aptima and LDT did not have any invalid or inhibited results.

Conclusion: The results indicate that the Aptima HSV-1 & 2 assay is sensitive and the performance characteristics of the Aptima assay is comparable to the assays analyzed for the detection and differentiation of HSV-1 and 2 from cutaneous and mucocutaneous lesions.

Keywords: Aptima; Automated; Cutaneous lesion; HSV; Mucocutaneous lesion; Performance evaluation.

PubMed Disclaimer

Publication types

MeSH terms

LinkOut - more resources