. 2018 Jun;20(3):368-377.
doi: 10.1007/s11307-017-1152-0.

Cytotoxicity Test Based on Human Cells Labeled with Fluorescent Proteins: Fluorimetry, Photography, and Scanning for High-Throughput Assay

Affiliations

Cytotoxicity Test Based on Human Cells Labeled with Fluorescent Proteins: Fluorimetry, Photography, and Scanning for High-Throughput Assay

Marina A Kalinina et al. Mol Imaging Biol. 2018 Jun.

Abstract

Purpose: High- and medium-throughput assays are now routine methods for drug screening and toxicology investigations on mammalian cells. However, a simple and cost-effective analysis of cytotoxicity that can be carried out with commonly used laboratory equipment is still required.

Procedures: The developed cytotoxicity assays are based on human cell lines stably expressing eGFP, tdTomato, mCherry, or Katushka2S fluorescent proteins. Red fluorescent proteins exhibit a higher signal-to-noise ratio, due to less interference by medium autofluorescence, in comparison to green fluorescent protein. Measurements have been performed on a fluorescence scanner, a plate fluorimeter, and a camera photodocumentation system.

Results: For a 96-well plate assay, the sensitivity per well and the measurement duration were 250 cells and 15 min for the scanner, 500 cells and 2 min for the plate fluorimeter, and 1000 cells and less than 1 min for the camera detection. These sensitivities are similar to commonly used MTT (tetrazolium dye) assays. The used scanner and the camera had not been previously applied for cytotoxicity evaluation. An image processing scheme for the high-resolution scanner is proposed that significantly diminishes the number of control wells, even for a library containing fluorescent substances. The suggested cytotoxicity assay has been verified by measurements of the cytotoxicity of several well-known cytotoxic drugs and further applied to test a set of novel bacteriotoxic compounds in a medium-throughput format.

Conclusion: The fluorescent signal of living cells is detected without disturbing them and adding any reagents, thus allowing to investigate time-dependent cytotoxicity effects on the same sample of cells. A fast, simple and cost-effective assay is suggested for cytotoxicity evaluation based on mammalian cells expressing fluorescent proteins and commonly used laboratory equipment.

Keywords: Cytotoxicity; Fluorescence detection; Fluorescent proteins; High throughput; Toxicity assay.

PubMed Disclaimer

References

    1. Biomed Khim. 2015 Nov-Dec;61(6):785-90 - PubMed
    1. Drug Discov Today. 2006 Dec;11(23-24):1054-60 - PubMed
    1. BMC Biotechnol. 2010 Jun 17;10:43 - PubMed
    1. Assay Drug Dev Technol. 2009 Jun;7(3):233-49 - PubMed
    1. Antimicrob Agents Chemother. 2016 Nov 21;60(12 ):7481-7489 - PubMed

Publication types

LinkOut - more resources