Mechanical Circulatory Support for the Failing Fontan: Conversion to Assisted Single Ventricle Circulation-Preliminary Observations

Affiliations

Mechanical Circulatory Support for the Failing Fontan: Conversion to Assisted Single Ventricle Circulation-Preliminary Observations

Syed M Peer et al. World J Pediatr Congenit Heart Surg. 2018 Jan.

Abstract

Background: Mechanical circulatory support (MCS) of a failing Fontan circulation remains challenging. We hypothesized that MCS can be provided by converting the Fontan circulation into a mechanically assisted single ventricle parallel circulation (MASVC).

Methods: A porcine model of functionally univentricular circulation was created under cardiopulmonary bypass (CPB) by performing an atrial septectomy, tricuspid valvectomy, and interrupting antegrade pulmonary blood flow. A centrifugal flow pump was placed with inflow from the common atrium. Eight millimeter Dacron grafts anastomosed to the ascending aorta and main pulmonary artery supplied systemic (Qs) and pulmonary (Qp) blood flow. Ultrasonic flow probes were used to measure Qs and Qp after weaning from CPB. The Qp/Qs ratio was regulated using an adjustable clamp. Hemodynamic and laboratory data were recorded.

Results: All four animals were successfully weaned from CPB onto the MASVC for a duration of two hours. Mechanically assisted single ventricle parallel circulation achieved satisfactory hemodynamics. As anticipated, the arterial oxygen saturation and partial pressure of oxygen in arterial blood were lower in the MASVC compared to baseline biventricular circulation. At the conclusion of the study, there was a trend towards a decrease in the mixed venous saturation with increasing oxygen extraction compared to the baseline. Serum lactate levels increased after weaning from CPB and did not return to baseline after two hours of support.

Conclusion: Mechanically assisted single ventricle parallel circulation can be established in a single ventricle animal model. This strategy could potentially provide MCS of a single ventricle circulation. Studies with longer duration of support are required to assess adequacy of support and long-term sustainability.

Keywords: BVAD; CHD; Fontan; RVAD; TAH); circulatory assist devices (LVAD; congenital heart disease (CHD); heart failure; univentricular heart.