Analytical and pre-analytical performance characteristics of a novel cartridge-type blood gas analyzer for point-of-care and laboratory testing
- PMID: 29339078
- DOI: 10.1016/j.clinbiochem.2018.01.007
Analytical and pre-analytical performance characteristics of a novel cartridge-type blood gas analyzer for point-of-care and laboratory testing
Abstract
Background: Point-of-care blood gas test results may benefit therapeutic decision making by their immediate impact on patient care. We evaluated the (pre-)analytical performance of a novel cartridge-type blood gas analyzer, the GEM Premier 5000 (Werfen), for the determination of pH, partial carbon dioxide pressure (pCO2), partial oxygen pressure (pO2), sodium (Na+), potassium (K+), chloride (Cl-), ionized calcium (iCa2+), glucose, lactate, and total hemoglobin (tHb).
Methods: Total imprecision was estimated according to the CLSI EP5-A2 protocol. The estimated total error was calculated based on the mean of the range claimed by the manufacturer. Based on the CLSI EP9-A2 evaluation protocol, a method comparison with the Siemens RapidPoint 500 and Abbott i-STAT CG8+ was performed. Obtained data were compared against preset quality specifications. Interference of potential pre-analytical confounders on co-oximetry and electrolyte concentrations were studied.
Results: The analytical performance was acceptable for all parameters tested. Method comparison demonstrated good agreement to the RapidPoint 500 and i-STAT CG8+, except for some parameters (RapidPoint 500: pCO2, K+, lactate and tHb; i-STAT CG8+: pO2, Na+, iCa2+ and tHb) for which significant differences between analyzers were recorded. No interference of lipemia or methylene blue on CO-oximetry results was found. On the contrary, significant interference for benzalkonium and hemolysis on electrolyte measurements were found, for which the user is notified by an interferent specific flag.
Conclusion: Identification of sample errors from pre-analytical sources, such as interferences and automatic corrective actions, along with the analytical performance, ease of use and low maintenance time of the instrument, makes the evaluated instrument a suitable blood gas analyzer for both POCT and laboratory use.
Keywords: Analytical quality specifications; Blood gas analyses; Imprecision; Method comparison; Point-of-care.
Copyright © 2018 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
Comment in
-
Novel generations of laboratory instruments should not worsen analytical quality: The case of GEM Premier 5000.Clin Biochem. 2018 Aug;58:128-130. doi: 10.1016/j.clinbiochem.2018.05.012. Epub 2018 May 22. Clin Biochem. 2018. PMID: 29792831 No abstract available.
Similar articles
-
Analytical performance evaluation of four cartridge-type blood gas analyzers.Clin Chem Lab Med. 2012 Feb 1;50(6):1083-91. doi: 10.1515/cclm-2011-0685. Clin Chem Lab Med. 2012. PMID: 22706251
-
Analytical performance evaluation of the new GEM® Premier™ 5000 analyzer in comparison to the GEM® Premier™ 4000 and the RapidPoint® 405 systems.Clin Chim Acta. 2018 Nov;486:313-319. doi: 10.1016/j.cca.2018.08.019. Epub 2018 Aug 15. Clin Chim Acta. 2018. PMID: 30118673
-
Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers.Ann Lab Med. 2022 Jul 1;42(4):467-472. doi: 10.3343/alm.2022.42.4.467. Ann Lab Med. 2022. PMID: 35177567 Free PMC article.
-
The Pre-Analytical Aspects of Blood Gas Analysis Point-Of-Care-Testing (POCT) and Their Impact on Clinical Decision-Making.Clin Lab. 2025 May 1;71(5). doi: 10.7754/Clin.Lab.2024.241029. Clin Lab. 2025. PMID: 40387753 Review.
-
Compatibility levels between blood gas analysis and central laboratory hemoglobin and electrolyte tests in pediatric patients: A single-center experience.Paediatr Anaesth. 2023 Feb;33(2):107-113. doi: 10.1111/pan.14567. Epub 2022 Oct 10. Paediatr Anaesth. 2023. PMID: 36178754 Review.
Cited by
-
Clinical performance validation of an integrated cartridge based bedside blood gas analyzer system in the acute care setting.Sci Rep. 2025 Jul 1;15(1):20620. doi: 10.1038/s41598-025-06710-6. Sci Rep. 2025. PMID: 40594851 Free PMC article.
-
Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers.Pract Lab Med. 2021 Apr 27;25:e00231. doi: 10.1016/j.plabm.2021.e00231. eCollection 2021 May. Pract Lab Med. 2021. PMID: 34095416 Free PMC article.
-
Lab at home: a promising prospect for on-site chemical and biological analysis.Anal Bioanal Chem. 2023 Jan;415(1):17-25. doi: 10.1007/s00216-022-04392-x. Epub 2022 Nov 5. Anal Bioanal Chem. 2023. PMID: 36334114 Free PMC article.
-
Recent advancements in critical care: Part I.Indian J Anaesth. 2023 Jan;67(1):102-109. doi: 10.4103/ija.ija_1005_22. Epub 2023 Jan 21. Indian J Anaesth. 2023. PMID: 36970474 Free PMC article.
-
Developing an animal model to detect drug-drug interactions impacting drug-induced respiratory depression.Toxicol Rep. 2020 Jan 25;7:188-197. doi: 10.1016/j.toxrep.2020.01.008. eCollection 2020. Toxicol Rep. 2020. PMID: 32021808 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials