Impact of missing attenuation and scatter corrections on 99m Tc-MAA SPECT 3D dosimetry for liver radioembolization using the patient relative calibration methodology: A retrospective investigation on clinical images
- PMID: 29383733
- DOI: 10.1002/mp.12774
Impact of missing attenuation and scatter corrections on 99m Tc-MAA SPECT 3D dosimetry for liver radioembolization using the patient relative calibration methodology: A retrospective investigation on clinical images
Abstract
Purpose: To investigate the clinical implication of performing pre-treatment dosimetry for 90 Y-microspheres liver radioembolization on 99m Tc-MAA SPECT images reconstructed without attenuation or scatter correction and quantified with the patient relative calibration methodology.
Methods: Twenty-five patients treated with SIR-Spheres® at Istituto Europeo di Oncologia and 31 patients treated with TheraSphere® at Istituto Nazionale Tumori were considered. For each acquired 99m Tc-MAA SPECT, four reconstructions were performed: with attenuation and scatter correction (AC_SC), only attenuation (AC_NoSC), only scatter (NoAC_SC) and without corrections (NoAC_NoSC). Absorbed dose maps were calculated from the activity maps, quantified applying the patient relative calibration to the SPECT images. Whole Liver (WL) and Tumor (T) regions were drawn on CT images. Injected Liver (IL) region was defined including the voxels receiving absorbed dose >3.8 Gy/GBq. Whole Healthy Liver (WHL) and Healthy Injected Liver (HIL) regions were obtained as WHL = WL - T and HIL = IL - T. Average absorbed dose to WHL and HIL were calculated, and the injection activity was derived following each Institute's procedure. The values obtained from AC_NoSC, NoAC_SC and NoAC_NoSC images were compared to the reference value suggested by AC_SC images using Bland-Altman analysis and Wilcoxon paired test (5% significance threshold). Absorbed-dose maps were compared to the reference map (AC_SC) in global terms using the Voxel Normalized Mean Square Error (%VNMSE), and at voxel level by calculating for each voxel the normalized difference with the reference value. The uncertainty affecting absorbed dose at voxel level was accounted for in the comparison; to this purpose, the voxel counts fluctuation due to Poisson and reconstruction noise was estimated from SPECT images of a water phantom acquired and reconstructed as patient images.
Results: NoAC_SC images lead to activity prescriptions not significantly different from the reference AC_SC images; the individual differences (<0.1 GBq for all IEO patients, <0.6 GBq for all but one INT patients) were comparable to the uncertainty affecting activity measurement. AC_NoSC and NoAC_NoSC images, instead, yielded significantly different activity prescriptions and wider 95% confidence intervals in the Bland-Altman analysis. Concerning the absorbed dose map, AC_NoSC images had the smallest %VNMSE value and the highest fraction of voxels differing less than 2 standard deviations from AC_SC.
Conclusions: The patient relative calibration methodology can compensate for the missing attenuation correction when performing healthy liver pre-treatment dosimetry: safe treatments can be planned even on NoAC_SC images, suggesting activities comparable to AC_SC images. Scatter correction is recommended due to its heavy impact on healthy liver dosimetry.
Keywords: SPECT corrections; SPECT quantification; internal dosimetry accuracy; liver radioembolization; voxel dosimetry.
© 2018 American Association of Physicists in Medicine.
Similar articles
-
Simplification of dosimetry in 90Y-radioembolization therapy by dual planar images.BMC Cancer. 2022 Dec 8;22(1):1288. doi: 10.1186/s12885-022-10392-y. BMC Cancer. 2022. PMID: 36482312 Free PMC article.
-
Impact of SPECT corrections on 3D-dosimetry for liver transarterial radioembolization using the patient relative calibration methodology.Med Phys. 2016 Jul;43(7):4053. doi: 10.1118/1.4953203. Med Phys. 2016. PMID: 27370124
-
Prognostic utility of 90Y radioembolization dosimetry based on fusion 99mTc-macroaggregated albumin-99mTc-sulfur colloid SPECT.J Nucl Med. 2013 Dec;54(12):2055-61. doi: 10.2967/jnumed.113.123257. Epub 2013 Oct 21. J Nucl Med. 2013. PMID: 24144563
-
Clinical impact of (99m)Tc-MAA SPECT/CT-based dosimetry in the radioembolization of liver malignancies with (90)Y-loaded microspheres.Eur J Nucl Med Mol Imaging. 2016 Mar;43(3):559-75. doi: 10.1007/s00259-015-3157-8. Epub 2015 Sep 4. Eur J Nucl Med Mol Imaging. 2016. PMID: 26338177 Free PMC article. Review.
-
The impact of PET and SPECT on dosimetry for targeted radionuclide therapy.Z Med Phys. 2006;16(1):47-59. doi: 10.1078/0939-3889-00291. Z Med Phys. 2006. PMID: 16696370 Review.
Cited by
-
EANM dosimetry committee series on standard operational procedures: a unified methodology for 99mTc-MAA pre- and 90Y peri-therapy dosimetry in liver radioembolization with 90Y microspheres.EJNMMI Phys. 2021 Nov 12;8(1):77. doi: 10.1186/s40658-021-00394-3. EJNMMI Phys. 2021. PMID: 34767102 Free PMC article.
-
166Ho microsphere scout dose for more accurate radioembolization treatment planning.Eur J Nucl Med Mol Imaging. 2020 Apr;47(4):744-747. doi: 10.1007/s00259-019-04617-9. Eur J Nucl Med Mol Imaging. 2020. PMID: 31875243 No abstract available.
-
An international phantom study of inter-site variability in Technetium-99m image quantification: analyses from the TARGET radioembolization study.EJNMMI Phys. 2024 May 29;11(1):46. doi: 10.1186/s40658-024-00647-x. EJNMMI Phys. 2024. PMID: 38809320 Free PMC article.
-
Simplification of dosimetry in 90Y-radioembolization therapy by dual planar images.BMC Cancer. 2022 Dec 8;22(1):1288. doi: 10.1186/s12885-022-10392-y. BMC Cancer. 2022. PMID: 36482312 Free PMC article.
-
Predictive value of 99mTc-MAA-based dosimetry in personalized 90Y-SIRT planning for liver malignancies.EJNMMI Res. 2023 Jul 3;13(1):63. doi: 10.1186/s13550-023-01011-3. EJNMMI Res. 2023. PMID: 37395912 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources