. 2018 Mar;36(3):479-487.
doi: 10.1097/HJH.0000000000001635.

Improving the accuracy of blood pressure measurement: the influence of the European Society of Hypertension International Protocol (ESH-IP) for the validation of blood pressure measuring devices and future perspectives

Affiliations

Improving the accuracy of blood pressure measurement: the influence of the European Society of Hypertension International Protocol (ESH-IP) for the validation of blood pressure measuring devices and future perspectives

George S Stergiou et al. J Hypertens. 2018 Mar.

Abstract

Introduction: The European Society of Hypertension (ESH) International Protocol (ESH-IP) for the validation of blood pressure (BP) measuring devices was published in 2002, with the main objective of simplifying the validation procedures, so that more BP monitors would be subjected to independent validation. This article provides an overview of the international impact of the ESH-IP and of the lessons learned from its use, to be able to justify further developments in validation protocols.

Methods: A review of published (PubMed) validation studies from 2002 to 2017 was performed.

Results: One hundred and seventy-seven validation studies using the ESH-IP, 59 using the British Hypertension Society protocol, 46 using the Association for the Advancement of Medical Instrumentation (AAMI) standard and 23 using the International Organization for Standardization (ISO) standard were identified. Lists of validated office-clinic, home and ambulatory BP monitors are provided. Of the ESH-IP studies, 93% tested oscillometric devices, 80% upper arm, 71% home, 25% office and 7% ambulatory monitors (some had more than one function).

Conclusion: The original goal of the ESH-IP has been fulfilled in that in the last decade the number of published validation studies has more than doubled. It is now recognized that the provision of accurate devices would be best served by having a universal protocol. An international initiative has been put in place by AAMI, ESH and ISO experts aiming to reach consensus for a universal validation protocol to be accepted worldwide, which will allow a more thorough evaluation of the accuracy and performance of future BP monitors.