. 2018 Mar 26;56(4):e01672-17.
doi: 10.1128/JCM.01672-17. Print 2018 Apr.

Evaluation of the Accelerate Pheno System: Results from Two Academic Medical Centers

Affiliations

Evaluation of the Accelerate Pheno System: Results from Two Academic Medical Centers

Joseph D Lutgring et al. J Clin Microbiol. .

Abstract

Rapid diagnostic tests are needed to improve patient care and to combat the problem of antimicrobial resistance. The Accelerate Pheno system (Accelerate Diagnostics, Tucson, AZ) is a new diagnostic device that can provide rapid bacterial identification and antimicrobial susceptibility test (AST) results directly from a positive blood culture. The device was compared to the standard of care at two academic medical centers. There were 298 blood cultures included in the study, and the Accelerate Pheno system provided a definitive identification result in 218 instances (73.2%). The Accelerate Pheno system provided a definitive and correct result for 173 runs (58.1%). The Accelerate Pheno system demonstrated an overall sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of 94.7%, 98.9%, 83.7%, and 99.7%, respectively. An AST result was available for analysis in 146 instances. The overall category agreement was 94.1% with 12 very major errors, 5 major errors, and 55 minor errors. After a discrepancy analysis, there were 5 very major errors and 4 major errors. The Accelerate Pheno system provided an identification result in 1.4 h and an AST result in 6.6 h; the identification and AST results were 41.5 h and 48.4 h faster than those with the standard of care, respectively. This study demonstrated that the Accelerate Pheno system is able to provide fast and accurate organism identification and AST data. A limitation is the frequency with which cultures required the use of alternative identification and AST methods.

Keywords: antimicrobial resistance; antimicrobial susceptibility testing; bacteremia; bloodstream infections; diagnostics; rapid diagnostic testing.

PubMed Disclaimer

Figures

FIG 1
FIG 1
Flowchart for isolates available for identification and AST analysis.

References

    1. Angus DC. 2011. Management of sepsis: a 47-year-old woman with an indwelling intravenous catheter and sepsis. JAMA 305:1469–1477. doi:10.1001/jama.2011.438. - DOI - PubMed
    1. Annane D, Bellissant E, Cavaillon JM. 2005. Septic shock. Lancet 365:63–78. doi:10.1016/S0140-6736(04)17667-8. - DOI - PubMed
    1. Hotchkiss RS, Karl IE. 2003. The pathophysiology and treatment of sepsis. N Engl J Med 348:138–150. doi:10.1056/NEJMra021333. - DOI - PubMed
    1. Singer M, Deutschman CS, Seymour CW, Shankar-Hari M, Annane D, Bauer M, Bellomo R, Bernard GR, Chiche JD, Coopersmith CM, Hotchkiss RS, Levy MM, Marshall JC, Martin GS, Opal SM, Rubenfeld GD, van der Poll T, Vincent JL, Angus DC. 2016. The third international consensus definitions for sepsis and septic shock (Sepsis-3). JAMA 315:801–810. doi:10.1001/jama.2016.0287. - DOI - PMC - PubMed
    1. Ferrer R, Martin-Loeches I, Phillips G, Osborn TM, Townsend S, Dellinger RP, Artigas A, Schorr C, Levy MM. 2014. Empiric antibiotic treatment reduces mortality in severe sepsis and septic shock from the first hour: results from a guideline-based performance improvement program. Crit Care Med 42:1749–1755. doi:10.1097/CCM.0000000000000330. - DOI - PubMed

Publication types

MeSH terms

Substances

LinkOut - more resources