Dual-energy CT: a phantom comparison of different platforms for abdominal imaging
- PMID: 29404773
- DOI: 10.1007/s00330-017-5238-5
Dual-energy CT: a phantom comparison of different platforms for abdominal imaging
Abstract
Objectives: Evaluation of imaging performance across dual-energy CT (DECT) platforms, including dual-layer CT (DLCT), rapid-kVp-switching CT (KVSCT) and dual-source CT (DSCT).
Methods: A semi-anthropomorphic abdomen phantom was imaged on these DECT systems. Scans were repeated three times for CTDIvol levels of 10 mGy, 20 mGy, 30 mGy and different fat-simulating extension rings. Over the available range of virtual-monoenergetic images (VMI), noise as well as quantitative accuracy of hounsfield units (HU) and iodine concentrations were evaluated.
Results: For all VMI levels, HU values could be determined with high accuracy compared to theoretical values. For KVSCT and DSCT, a noise increase was observed towards lower VMI levels. A patient-size dependent increase in the uncertainty of quantitative iodine concentrations is observed for all platforms. For a medium patient size the iodine concentration root-mean-square deviation at 20 mGy is 0.17 mg/ml (DLCT), 0.30 mg/ml (KVSCT) and 0.77mg/ml (DSCT).
Conclusion: Noticeable performance differences are observed between investigated DECT systems. Iodine concentrations and VMI HUs are accurately determined across all DECT systems. KVSCT and DLCT deliver slightly more accurate iodine concentration values than DSCT for investigated scenarios. In DLCT, low-noise and high-image contrast at low VMI levels may help to increase diagnostic information in abdominal CT.
Key points: • Current dual-energy CT platforms provide accurate, reliable quantitative information. • Dual-energy CT cross-platform evaluation revealed noticeable performance differences between different systems. • Dual-layer CT offers constant noise levels over the complete energy range.
Keywords: Comparative study; Computed tomography, X-ray; Iodine; Quantitative evaluation; Radiologic phantom.
Similar articles
-
Performance of four dual-energy CT platforms for abdominal imaging: a task-based image quality assessment based on phantom data.Eur Radiol. 2021 Jul;31(7):5324-5334. doi: 10.1007/s00330-020-07671-2. Epub 2021 Jan 15. Eur Radiol. 2021. PMID: 33449188
-
Comparison of virtual monoenergetic imaging between a rapid kilovoltage switching dual-energy computed tomography with deep-learning and four dual-energy CTs with iterative reconstruction.Quant Imaging Med Surg. 2022 Feb;12(2):1149-1162. doi: 10.21037/qims-21-708. Quant Imaging Med Surg. 2022. PMID: 35111612 Free PMC article.
-
Assessment of quantification accuracy and image quality of a full-body dual-layer spectral CT system.J Appl Clin Med Phys. 2018 Jan;19(1):204-217. doi: 10.1002/acm2.12243. Epub 2017 Dec 20. J Appl Clin Med Phys. 2018. PMID: 29266724 Free PMC article.
-
Dual-energy CT revisited with multidetector CT: review of principles and clinical applications.Diagn Interv Radiol. 2011 Sep;17(3):181-94. doi: 10.4261/1305-3825.DIR.3860-10.0. Epub 2010 Nov 14. Diagn Interv Radiol. 2011. PMID: 20945292 Review.
-
Dual-energy CT applications in the abdomen.AJR Am J Roentgenol. 2012 Nov;199(5 Suppl):S64-70. doi: 10.2214/AJR.12.9196. AJR Am J Roentgenol. 2012. PMID: 23097169 Review.
Cited by
-
Determining the Optimal Energy Level of Virtual Monoenergetic Images in Dual-Source CT for Diagnosis of Bowel Obstruction and Colitis.Diagnostics (Basel). 2023 Nov 21;13(23):3491. doi: 10.3390/diagnostics13233491. Diagnostics (Basel). 2023. PMID: 38066732 Free PMC article.
-
Five material tissue decomposition by dual energy computed tomography.Sci Rep. 2022 Oct 12;12(1):17117. doi: 10.1038/s41598-022-21193-5. Sci Rep. 2022. PMID: 36224229 Free PMC article.
-
First-generation clinical dual-source photon-counting CT: ultra-low-dose quantitative spectral imaging.Eur Radiol. 2022 Dec;32(12):8579-8587. doi: 10.1007/s00330-022-08933-x. Epub 2022 Jun 16. Eur Radiol. 2022. PMID: 35708838 Free PMC article.
-
Reply to the Letter to the Editor: Quantitative accuracy of virtual monoenergetic images from multi-energy CT.Eur Radiol. 2024 May;34(5):2960-2962. doi: 10.1007/s00330-023-10286-y. Epub 2023 Nov 7. Eur Radiol. 2024. PMID: 37934244 Free PMC article. No abstract available.
-
Liquid Embolic Agents in Spectral X-Ray Photon-Counting Computed Tomography using Tantalum K-Edge Imaging.Sci Rep. 2019 Mar 27;9(1):5268. doi: 10.1038/s41598-019-41737-6. Sci Rep. 2019. PMID: 30918297 Free PMC article.
References
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical