Normal and abnormal prosthetic valve function as assessed by Doppler echocardiography
- PMID: 2942593
- DOI: 10.1016/s0735-1097(86)80046-8
Normal and abnormal prosthetic valve function as assessed by Doppler echocardiography
Abstract
Doppler echocardiography was performed in 136 patients with a normally functioning prosthetic valve in the aortic (n = 59), mitral (n = 74) and tricuspid (n = 3) positions. These included patients with St. Jude (n = 82), Björk-Shiley (n = 18), Beall (n = 13), Starr-Edwards (n = 7) or tissue (n = 16) valves. Peak and mean pressure gradients across the prostheses were measured using the simplified Bernoulli equation. The prosthetic valve orifice (PVO, in square centimeters), only in the mitral position, was calculated by the equation: PVO = 220/pressure half-time. In the aortic position, the St. Jude valve had a lower peak velocity (2.3 +/- 0.6 m/s, range 1.0 to 3.9), peak gradient (22 +/- 12 mm Hg, range 4 to 61) and mean gradient (12 +/- 7 mm Hg, range 2 to 32) than the other valves (p less than 0.05) when compared with Starr-Edwards). In the mitral position, the St. Jude valve had the largest orifice (3.0 +/- 0.6 cm2, range 1.8 to 5.0) (p less than 0.0001 compared with all other valves). Insignificant regurgitation was commonly found by pulsed mode Doppler technique in patients with a St. Jude or Björk-Shiley valve in the aortic or mitral position and in patients with a Starr-Edwards or tissue valve in the aortic position. In 17 other patients with a malfunctioning prosthesis (four St. Jude, two Björk-Shiley, four Beall and seven tissue valves) proven by cardiac catheterization, surgery or autopsy, Doppler echocardiography correctly identified the complication (significant regurgitation or obstruction) in all but 2 patients who had a Beall valve. It is concluded that 1) the St. Jude valve appears to have the most optimal hemodynamics; mild regurgitation can be detected by the Doppler technique in normally functioning St. Jude and Björk-Shiley valves in the aortic or mitral position and in Starr-Edwards and tissue valves in the aortic position, and 2) Doppler echocardiography is a useful method for the detection of prosthetic valve malfunction, especially when the St. Jude, Björk-Shiley and tissue valves are assessed.
Similar articles
-
[Doppler echocardiography in assessing mechanical and biological heart valve prostheses].G Ital Cardiol. 1988 Feb;18(2):121-34. G Ital Cardiol. 1988. PMID: 3410201 Italian.
-
Doppler hemodynamic evaluation of prosthetic (Starr-Edwards and Björk-Shiley) and bioprosthetic (Hancock and Carpentier-Edwards) cardiac valves.Am J Cardiol. 1985 Aug 1;56(4):325-32. doi: 10.1016/0002-9149(85)90858-6. Am J Cardiol. 1985. PMID: 4025173
-
Analysis of Doppler-obtained velocity curves in functional evaluation of mechanical prosthetic valves in the mitral and aortic positions.J Am Soc Echocardiogr. 1988 May-Jun;1(3):211-25. doi: 10.1016/s0894-7317(88)80077-4. J Am Soc Echocardiogr. 1988. PMID: 3272769
-
Doppler echocardiography assessment of prosthetic heart valves.Echocardiography. 1990 Mar;7(2):97-114. doi: 10.1111/j.1540-8175.1990.tb00353.x. Echocardiography. 1990. PMID: 10149195 Review.
-
Current status of the Beall, Bjork-Shiley, Braunwald-Cutter, Lillehei-Kaster and Smeloff-Cutter cardiac valve prostheses.Am J Cardiol. 1975 Jun;35(6):855-65. doi: 10.1016/0002-9149(75)90122-8. Am J Cardiol. 1975. PMID: 124127 Review.
Cited by
-
Effect of respiration on Doppler parameters of normal tricuspid porcine bioprosthetic valves.Int J Card Imaging. 1995 Mar;11(1):55-8. doi: 10.1007/BF01148954. Int J Card Imaging. 1995. PMID: 7730682
-
Management of prosthetic heart valve complications.Curr Treat Options Cardiovasc Med. 2012 Dec;14(6):608-21. doi: 10.1007/s11936-012-0212-7. Curr Treat Options Cardiovasc Med. 2012. PMID: 23086156
-
Early Doppler-echocardiography evaluation of Carpentier-Edwards Standard and Carpentier-Edwards Magna aortic prosthetic valve: comparison of hemodynamic performance.Cardiovasc Ultrasound. 2011 Nov 24;9:37. doi: 10.1186/1476-7120-9-37. Cardiovasc Ultrasound. 2011. PMID: 22114985 Free PMC article.
-
Doppler echocardiographic characteristics of normal and dysfunctioning prosthetic valves in the tricuspid and mitral position.Br Heart J. 1990 Jan;63(1):41-4. doi: 10.1136/hrt.63.1.41. Br Heart J. 1990. PMID: 2310643 Free PMC article.
-
Effect of changes in heart rate on pressure half time in normally functioning mitral valve prostheses.Br Heart J. 1988 Dec;60(6):502-6. doi: 10.1136/hrt.60.6.502. Br Heart J. 1988. PMID: 3224053 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical