. 2018 Apr:54:106-111.
doi: 10.1016/j.clinbiochem.2018.02.005. Epub 2018 Feb 9.

Paper spray mass spectrometry for the direct, semi-quantitative measurement of fentanyl and norfentanyl in complex matrices

Affiliations

Paper spray mass spectrometry for the direct, semi-quantitative measurement of fentanyl and norfentanyl in complex matrices

Gregory W Vandergrift et al. Clin Biochem. 2018 Apr.

Abstract

Background: Fentanyl is a potent, synthetic opioid at the centre of an international health crisis that has seen thousands of fatal overdoses. Most analytical methods focus on the detection of fentanyl in blood and/or urine (i.e., post-drug use). Harm reduction (including pre-screening before drug use) may be an effective strategy against fentanyl overdoses.

Method: Paper spray-mass spectrometry (PS-MS) is an inexpensive, direct sampling strategy where a small volume of sample (<10 μL) is spotted onto a piece of paper that is then wetted and connected to high voltage. Ions are emitted from the paper and enter a mass spectrometer for sensitive and selective semi-quantitation using labeled internal standards.

Results: We present the use of PS-MS for the direct measurement of fentanyl and norfentanyl using a custom PS interface, demonstrating that paper tip position and quality can significantly affect quantitative results. Furthermore, we observe comparable calibrations for fentanyl and norfentanyl (0.5 to 600 ng/mL) across a variety of complex matrices (methanol, diluted urine, analgesic slurry). Detection limits for fentanyl are as low as 0.049 ng/mL (0.4 pg total material) in methanol, and 0.66 ng/mL (5.3 pg total material) spiked in an analgesic slurry (illicit substance simulation). PS-MS was compared with liquid chromatography-MS for the analyses of real urine samples, with satisfactory results.

Conclusion: PS-MS shows potential as a sensitive and selective direct measurement strategy for use in fentanyl harm reduction strategies, and may also be used for pre-screening in advance of or in combination with more conventional (i.e., chromatographic) analyses.

Keywords: Ambient ionization; Fentanyl; Norfentanyl; Opioid harm reduction; Paper spray; Tandem mass spectrometry.