Turbulent shear stress measurements in the vicinity of aortic heart valve prostheses
- PMID: 2943742
- DOI: 10.1016/0021-9290(86)90020-5
Turbulent shear stress measurements in the vicinity of aortic heart valve prostheses
Abstract
A two dimensional laser Doppler anemometer system has been used to measure the turbulent shear fields in the immediate downstream vicinity of a variety of mechanical and bioprosthetic aortic heart valves. The measurements revealed that all the mechanical valves studied, created regions of elevated levels of turbulent shear stress during the major portion of systole. The tissue bioprostheses also created elevated levels of turbulence, but they were confined to narrow regions in the bulk of the flow field. The newer generation of bioprostheses create turbulent shear stresses which are considerably lower than those created by the older generation tissue valve designs. All the aortic valves studied (mechanical and tissue) create turbulent shear stress levels which are capable of causing sub-lethal and/or lethal damage to blood elements.
Similar articles
-
Two-dimensional color-mapping of turbulent shear stress distribution downstream of two aortic bioprosthetic valves in vitro.J Biomech. 1992 Apr;25(4):429-40. doi: 10.1016/0021-9290(92)90262-y. J Biomech. 1992. PMID: 1583021
-
In vitro pulsatile flow velocity and turbulent shear stress measurements in the vicinity of mechanical aortic heart valve prostheses.Life Support Syst. 1985 Oct-Dec;3(4):283-312. Life Support Syst. 1985. PMID: 4068753
-
Pulsatile flow velocity and shear stress measurements on the St. Jude bileaflet valve prosthesis.Scand J Thorac Cardiovasc Surg. 1986;20(1):15-28. doi: 10.3109/14017438609105910. Scand J Thorac Cardiovasc Surg. 1986. PMID: 2939558
-
In vitro velocity and turbulence measurements in the vicinity of three new mechanical aortic heart valve prostheses: Björk-Shiley Monostrut, Omni-Carbon, and Duromedics.J Thorac Cardiovasc Surg. 1988 May;95(5):929-39. J Thorac Cardiovasc Surg. 1988. PMID: 3361941
-
Porcine valves: Hancock and Carpentier-Edwards aortic prostheses.Semin Thorac Cardiovasc Surg. 1996 Jul;8(3):259-68. Semin Thorac Cardiovasc Surg. 1996. PMID: 8843517 Review.
Cited by
-
Measurements of steady flow through a bileaflet mechanical heart valve using stereoscopic PIV.Med Biol Eng Comput. 2011 Mar;49(3):325-35. doi: 10.1007/s11517-010-0705-z. Epub 2010 Nov 16. Med Biol Eng Comput. 2011. PMID: 21080093
-
Assessment of prosthetic aortic valve performance by magnetic resonance velocity imaging.MAGMA. 2000 Feb;10(1):18-26. doi: 10.1007/BF02613108. MAGMA. 2000. PMID: 10697222
-
Multiscale structure and function of the aortic valve apparatus.Physiol Rev. 2024 Oct 1;104(4):1487-1532. doi: 10.1152/physrev.00038.2022. Epub 2023 Sep 21. Physiol Rev. 2024. PMID: 37732828 Free PMC article. Review.
-
High-resolution fluid-structure interaction simulations of flow through a bi-leaflet mechanical heart valve in an anatomic aorta.Ann Biomed Eng. 2010 Feb;38(2):326-44. doi: 10.1007/s10439-009-9807-x. Epub 2009 Oct 6. Ann Biomed Eng. 2010. PMID: 19806458 Free PMC article.
-
Transcatheter aortic valves produce unphysiological flows which may contribute to thromboembolic events: An in-vitro study.J Biomech. 2016 Dec 8;49(16):4080-4089. doi: 10.1016/j.jbiomech.2016.10.050. Epub 2016 Nov 3. J Biomech. 2016. PMID: 27836502 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources