Flush-mounted hot film anemometer accuracy in pulsatile flow
- PMID: 2943937
- DOI: 10.1115/1.3138607
Flush-mounted hot film anemometer accuracy in pulsatile flow
Abstract
The accuracy of a flush-mounted hot film anemometer probe for wall shear stress measurements in physiological pulsatile flows was evaluated in fully developed pulsatile flow in a rigid straight tube. Measured wall shear stress waveform based on steady flow anemometer probe calibrations were compared to theoretical wall shear stress waveforms based on well-established theory and measured flow rate waveforms. The measured and theoretical waveforms were in close agreement during systole (average deviation of 14 percent at peak systole). As expected, agreement was poor during diastole because of flow reversal and diminished frequency response at low shear rate.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Research Materials