High-Resolution Infrared Thermography of Esophageal Temperature During Radiofrequency Ablation of Atrial Fibrillation

Affiliations

High-Resolution Infrared Thermography of Esophageal Temperature During Radiofrequency Ablation of Atrial Fibrillation

Matthew G Daly et al. Circ Arrhythm Electrophysiol. 2018 Feb.

Abstract

Background: Catheter ablation for atrial fibrillation has potential to cause esophageal thermal injury. Esophageal temperature monitoring during ablation is commonly used; however, it has not eliminated thermal injuries, possibly because conventional sensors have poor spatial sampling and response characteristics. To enhance understanding of temperature dynamics that may underlie esophageal injury, we tested a high-resolution, intrabody, infrared thermography catheter to continuously image esophageal temperatures during ablation.

Methods and results: Atrial fibrillation ablation patients were instrumented with a flexible, 9F infrared temperature catheter inserted nasally (n=8) or orally (n=8) into the esophagus adjacent to the left atrium. Ablation was performed while the infrared catheter continuously recorded surface temperatures from 7680 points per second circumferentially over a 6-cm length of esophagus. Physicians were blinded to temperature data. Endoscopy was performed within 24 hours to document esophageal injury. Thermal imaging showed that most patients (10/16) experienced ≥1 events where peak esophageal temperature was >40°C. Three patients experienced temperatures >50°C; and 1 experienced >60°C. Analysis of temperature data for each subject's maximum thermal event revealed high gradients (2.3±1.4°C/mm) and rates of change (1.5±1.3°C/s) with an average length of esophageal involvement of 11.0±5.4 mm. Endoscopy identified 3 distinct thermal lesions, all in patients with temperatures >50°C; all resolved within 2 weeks.

Conclusions: Infrared thermography provided dynamic, high-resolution mapping of esophageal temperatures during cardiac ablation. Esophageal thermal injury occurred with temperatures >50°C and was associated with large spatiotemporal gradients. Additional studies are warranted to determine the relationships between thermal parameters and esophageal injury.

Keywords: atrial fibrillation; catheter ablation; endoscopy; temperature; thermography.