Comparison of Corneal Tomography: Repeatability, Precision, Misalignment, Mean Elevation, and Mean Pachymetry
- PMID: 29482368
- DOI: 10.1080/02713683.2018.1441873
Comparison of Corneal Tomography: Repeatability, Precision, Misalignment, Mean Elevation, and Mean Pachymetry
Abstract
Purpose: To compare corneal tomography and its statistical uncertainty for measurements obtained by three clinically used corneal tomographers: A Scheimpflug camera (Pentacam HR), a swept source optical coherence tomography system (CASIA SS-1000), and Placido ring imaging (TMS-5).
Material and methods: Repeated measurements with all three devices on 34 normal eyes were used to estimate the repeatability, precision, and mean values of corneal elevation and pachymetry within 8 mm diameter. The repeatability (standard deviation) was calculated for each data point of the corneal elevation data-maps of anterior and posterior cornea as well as for the pachymetry data-maps. Uncertainty on the position of the eye at each measurement might contribute to the differences between elevation data-maps. To take this into account, we defined the precision as the standard deviation for the elevation data-maps of anterior and posterior cornea after correction of misalignment-effects (rotation, translation). The mean elevation and pachymetry data-maps were fitted with Zernike polynomials for interdevice-comparison.
Results: Pentacam HR offered the best repeatability and precision for the anterior corneal elevation (<3 and <1.6 μm, respectively). CASIA SS-1000 offered good repeatability and precision with high resolution for posterior corneal elevation, and the best repeatability for pachymetry (<3 μm). TMS-5 measured anterior elevation with similar repeatability to CASIA SS-1000 (<6 μm). The data-maps of the three tomographers could not be used interchangeably. The largest differences were observed for pachymetry and posterior corneal elevation data-maps.
Conclusions: Misalignment limited the repeatability of TMS-5 and Pentacam HR, but had little influence on the repeatability of CASIA SS-1000.
Keywords: Corneal shape; corneal diagnostics; corneal height; corneal thickness; interdevice comparison.
Similar articles
-
Comparison of corneal elevation and pachymetry measurements made by two state of the art corneal tomographers with different measurement principles.PLoS One. 2019 Oct 16;14(10):e0223770. doi: 10.1371/journal.pone.0223770. eCollection 2019. PLoS One. 2019. PMID: 31618270 Free PMC article.
-
A Concept for the analysis of repeatability and precision of corneal shape measurements.Z Med Phys. 2016 Jun;26(2):150-8. doi: 10.1016/j.zemedi.2016.01.002. Epub 2016 Feb 10. Z Med Phys. 2016. PMID: 26874793
-
Precision of a new Scheimpflug and Placido-disk analyzer in measuring corneal thickness and agreement with ultrasound pachymetry.J Cataract Refract Surg. 2013 Feb;39(2):219-24. doi: 10.1016/j.jcrs.2012.10.034. Epub 2012 Dec 3. J Cataract Refract Surg. 2013. PMID: 23218819
-
Measuring the cornea: the latest developments in corneal topography.Curr Opin Ophthalmol. 2007 Jul;18(4):325-33. doi: 10.1097/ICU.0b013e3281ca7121. Curr Opin Ophthalmol. 2007. PMID: 17568210 Review.
-
Cornea and anterior eye assessment with placido-disc keratoscopy, slit scanning evaluation topography and scheimpflug imaging tomography.Indian J Ophthalmol. 2018 Mar;66(3):360-366. doi: 10.4103/ijo.IJO_850_17. Indian J Ophthalmol. 2018. PMID: 29480244 Free PMC article. Review.
Cited by
-
Effects of intracorneal ring segments implementation technique and design on corneal biomechanics and keratometry in a personalized computational analysis.Sci Rep. 2021 Jul 13;11(1):14433. doi: 10.1038/s41598-021-93821-5. Sci Rep. 2021. PMID: 34257343 Free PMC article.
-
Repeatability of Corneal Densitometry Measurements using a Scheimpflug Camera in Healthy Normal Corneas.J Curr Ophthalmol. 2022 Apr 16;34(1):50-55. doi: 10.4103/joco.joco_173_21. eCollection 2022 Jan-Mar. J Curr Ophthalmol. 2022. PMID: 35620364 Free PMC article.
-
Comparison of a New Scheimpflug Camera and Swept-Source Optical Coherence Tomographer for Measurements of Anterior Segment Parameters.Ophthalmol Ther. 2023 Dec;12(6):3187-3198. doi: 10.1007/s40123-023-00813-x. Epub 2023 Sep 25. Ophthalmol Ther. 2023. PMID: 37747639 Free PMC article.
-
[Reliability of corneal tomography after implantation of intracorneal ring segments for keratoconus].Ophthalmologe. 2020 Nov;117(11):1092-1099. doi: 10.1007/s00347-020-01074-w. Ophthalmologe. 2020. PMID: 32140771 Free PMC article. German.
-
Ray tracing optimization: a new method for intraocular lens power calculation in regular and irregular corneas.Sci Rep. 2023 Mar 20;13(1):4555. doi: 10.1038/s41598-023-31525-8. Sci Rep. 2023. PMID: 36941337 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources