. 2018 May;19(3):351-354.
doi: 10.1002/acm2.12290. Epub 2018 Mar 1.

Validating kQ =1.0 assumption in TG51 with PTW 30013 farmer chamber for Varian TrueBeam's 2.5 MV imaging beam

Affiliations

Validating kQ =1.0 assumption in TG51 with PTW 30013 farmer chamber for Varian TrueBeam's 2.5 MV imaging beam

Shelby Grzetic et al. J Appl Clin Med Phys. 2018 May.

Abstract

AAPM Report 142 recommends and the State of Ohio requires that the imaging dose be quantified in radiotherapy applications. Using the TG51 dose calibration protocol for MV Imaging dose measurement requires knowledge of the kQ parameter for the beam quality and the ionization chamber type under investigation. The %dd(10)x of the Varian TrueBeam 2.5 MV imaging beam falls outside the range of the available data for the calculation of the kQ value. Due to the similarities of the 2.5 MV imaging beam and the 60 Co beam, we and others made the assumption that kQ = 1.0 in TG51 calculations. In this study, we used the TG21 and TG51 calibration protocols in conjunction to validate that kQ = 1.0 for the 2.5 MV imaging beam using a PTW 30013 farmer chamber. Standard measurements for TG51 absolute dosimetry QA were performed at 100 cm SSD, 10 cm depth, 10 × 10 field size, delivering 100 Monitor Units to a waterproof Farmer Chamber (PTW TN30013) for both 2.5 and 6 MV. Both the TG21 and TG51 formalisms were used to calculate the dose to water per MU at dmax (Dw /MU) for the 6 MV beam. The calculated outputs were 1.0005 and 1.0004 cGy/MU respectively. The TG21 formalism was then used to calculate (Dw /MU) for the 2.5 MV imaging beam. This value was then used in the TG51 formalism to find kQ for the 2.5 MV imaging beam. A kQ value of 1.00 ± 0.01 was calculated for 2.5 MV using this method.

Keywords: 2.5 MV imaging dose; dose calibration.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Measured PDD for the 2.5 MV imaging beam after shifting to the effective point of measurement.

References

    1. Klein EE, Hanley J, Bayouth J, et al. Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36:4197–4212. - PubMed
    1. Code OA. Radiation Control, Quality assurance for radiation therapy, simulation and image guidance equipment. 2016; 3701: 1‐67‐09, http://codes.ohio.gov/oac/3701:1-67-09
    1. Almond PR, Biggs PJ, Coursey BM, et al. AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beams. Med Phys. 1999;26:1847–1870. - PubMed
    1. Task Group RTCA . A protocol for the determination of absorbed dose from high‐energy photon and electron beams. Med Phys. 1983;10:741–771. - PubMed
    1. McEwen M, DeWerd L, Ibbott G, et al. Addendum to the AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon beams. Med Phys. 2014;41:041501. - PMC - PubMed