Consistency of electronic measurements of endodontic working length when using multiple devices from the same manufacturer-an in vitro study
- PMID: 29516188
- DOI: 10.1007/s00784-018-2404-3
Consistency of electronic measurements of endodontic working length when using multiple devices from the same manufacturer-an in vitro study
Abstract
Objectives: To assess the consistency of electronic determination of endodontic working length obtained from four identical electronic root canal length measurement devices (ERCLMD) from five different types of ERCLMD each under various in vitro conditions.
Materials and methods: Eight extracted teeth, seven single-rooted teeth, and one molar were accessed. Root length was measured and instruments were inserted and fixed with the tip placed beyond and short of the apical constriction, in roots with an artificial perforation or an open apex. Devices tested were Root ZX (Morita, Kyoto, Japan), Dentaport ZX (Morita), Apex ID (SybronEndo, Glendora, USA), ProPex II (Dentsply Maillefer, Victoria, Australia), and Raypex 6 (VDW-Antaeos, Munich, Germany). Teeth were irrigated with different solutions (NaOCl, EDTA, CHX). ERCLMDs were connected and measurements were recorded. Consistency was classified by the scores 0-4. Comparisons were carried out using the Kruskal-Wallis test (α = 0.05). For multiple testing, the level of significance was adjusted and analysis was performed using the Mann-Whitney U test.
Results: Among the five types of ERCLMD, Apex ID and Raypex 6 showed the highest consistency. There was no statistically significant difference between the settings and conditions. Raypex 6 showed the highest consistency for measurements in case of a perforation.
Conclusion: Different devices from one type of ERCLMD show a high consistency. Nevertheless, general statements on the accuracy of one type of ERCLMD taken from studies investigating only one device per type should be drawn with caution.
Clinical relevance: The study refers to the reliability and repeatability of determination of endodontic working length by using ERCLMDs.
Keywords: Electronic root canal length measurement device; Intra-device consistency; Open apex; Perforation; Working length.
Similar articles
-
Comparison of four electronic root canal length measurement devices.Acta Odontol Scand. 2017 Jul;75(5):325-331. doi: 10.1080/00016357.2017.1307447. Epub 2017 Mar 29. Acta Odontol Scand. 2017. PMID: 28355939
-
Influence of apical periodontitis on the accuracy of 3 electronic root canal length measurement devices: an in vivo study.J Endod. 2014 Mar;40(3):355-9. doi: 10.1016/j.joen.2013.11.012. Epub 2013 Dec 27. J Endod. 2014. PMID: 24565652
-
In vivo accuracy of three electronic root canal length measurement devices: Dentaport ZX, Raypex 5 and ProPex II.Int Endod J. 2012 Jun;45(6):552-6. doi: 10.1111/j.1365-2591.2011.02010.x. Epub 2012 Jan 19. Int Endod J. 2012. PMID: 22257341 Clinical Trial.
-
Influence of instrument size and varying electrical resistance of root canal instruments on accuracy of three electronic root canal length measurement devices.Int Endod J. 2017 May;50(5):506-511. doi: 10.1111/iej.12649. Epub 2016 May 5. Int Endod J. 2017. PMID: 27079789
-
[Electronic determination of the length of the root canal].Rev Belge Med Dent (1984). 1999;54(4):259-81. Rev Belge Med Dent (1984). 1999. PMID: 12723167 Review. French.
Cited by
-
The influence of periapical lesions on the repeatability of two electronic apex locators in vivo.Clin Oral Investig. 2021 Sep;25(9):5239-5245. doi: 10.1007/s00784-021-03831-w. Epub 2021 Feb 13. Clin Oral Investig. 2021. PMID: 33580848
-
Influence of the Maxillary Sinus on the Accuracy of the Root ZX Apex Locator: An Ex Vivo Study.Dent J (Basel). 2019 Jan 2;7(1):3. doi: 10.3390/dj7010003. Dent J (Basel). 2019. PMID: 30609762 Free PMC article.
-
Accuracy of different generations of apex locators in determining working length; a systematic review and meta-analysis.Saudi Dent J. 2022 Jan;34(1):11-20. doi: 10.1016/j.sdentj.2021.09.020. Epub 2021 Sep 25. Saudi Dent J. 2022. PMID: 35068894 Free PMC article. Review.
-
Evaluation of the accuracy of different apex locators in determining the working length during root canal retreatment.J Dent Res Dent Clin Dent Prospects. 2020 Spring;14(2):125-129. doi: 10.34172/joddd.2020.026. Epub 2020 Jun 17. J Dent Res Dent Clin Dent Prospects. 2020. PMID: 32908654 Free PMC article.
-
Influence of Different Apical Foramen Morphologies on the Accuracy of Four Electronic Foramen Locators.Eur J Dent. 2024 Oct;18(4):1101-1106. doi: 10.1055/s-0044-1782214. Epub 2024 May 2. Eur J Dent. 2024. PMID: 38698613 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources