Attempted salvage of infected cardiovascular implantable electronic devices: Are there clinical factors that predict success?

Affiliations
Observational Study

Attempted salvage of infected cardiovascular implantable electronic devices: Are there clinical factors that predict success?

James E Peacock Jr et al. Pacing Clin Electrophysiol. 2018 May.

Abstract

Background: Published guidelines mandate complete device removal in cases of cardiovascular implantable electronic device (CIED) infection. Clinical predictors of successful salvage of infected CIEDs have not been defined.

Methods: Data from the Multicenter Electrophysiologic Device Infection Collaboration, a prospective, observational, multinational cohort study of CIED infection, were used to investigate whether clinical predictors of successful salvage of infected devices could be identified.

Results: Of 433 adult patients with CIED infections, 306 (71%) underwent immediate device explantation. Medical management with device retention and antimicrobial therapy was initially attempted in 127 patients (29%). "Early failure" of attempted salvage occurred in 74 patients (58%) who subsequently underwent device explantation during the index hospitalization. The remaining 53 patients (42%) in the attempted salvage group retained their CIED. Twenty-six (49%) had resolution of CIED infection (successful salvage group) whereas 27 patients (51%) experienced "late" salvage failure. Upon comparing the salvage failure group, early and late (N = 101), to the group experiencing successful salvage of an infected CIED (N = 26), no clinical or laboratory predictors of successful salvage were identified. However, by univariate analysis, coagulase-negative staphylococci as infecting pathogens (P = 0.0439) and the presence of a lead vegetation (P = 0.024) were associated with overall failed salvage.

Conclusions: In patients with definite CIED infections, clinical and laboratory variables cannot predict successful device salvage. Until new data are forthcoming, device explantation should remain a mandatory and early management intervention in patients with CIED infection in keeping with existing expert guidelines unless medical contraindications exist or patients refuse device removal.

Keywords: ICD; cardiovascular implantable electronic device (CIED); infection; pacemaker; salvage.