Keratoprosthesis Decentration and Tilt Results in Degradation in Image Quality

Affiliations

Keratoprosthesis Decentration and Tilt Results in Degradation in Image Quality

Richard W Helms et al. Cornea. 2018 Jun.

Abstract

Purpose: To study the effect of decentration and tilt of the type I Boston keratoprosthesis (KPro) on image quality in both aphakic and pseudophakic eyes.

Methods: An optical ray-tracing program was used to simulate the image projected onto the retina in an eye with a perfectly centered KPro, and in eyes with varying degrees of KPro decentration and tilt. Decentration was modeled along a typical white-to-white distance of 12.0 mm, and the corresponding tilt was calculated assuming a radius of curvature of 8.0 mm, the radius of curvature of the backplate of the KPro. Both aphakic and pseudophakic eyes were simulated, and the corresponding modulation transfer function curves, point spread functions, and astigmatism were analyzed.

Results: The perfectly centered KPro produced a high-quality image with no induced astigmatism. Increasing decentration beyond approximately 0.5 mm resulted in poorer image quality with a more pronounced effect in the presence of an intraocular lens. Using models of the normal eye as a threshold, image degradation due to decentration becomes clinically significant at approximately 1.4 mm and 0.9 mm for the aphakic and pseudophakic cases, respectively. Astigmatism introduced by decentration is approximately 0.25 D cylinders at those thresholds.

Conclusions: Decentration of up to 0.5 mm had no significant impact on image quality and an attempt at good intraoperative centration of the KPro within this range is important. Conversely, decentration of 0.9 mm or more during surgical implantation can result in significant degradation in retinal image quality including astigmatism. The effect is greater in the pseudophakic eye.