Ultrasound-guided intravenous access in adults using SonoStik®, a novel encapsulated sterile guidewire: A prospective cohort trial
- PMID: 29529917
- DOI: 10.1177/1129729818758228
Ultrasound-guided intravenous access in adults using SonoStik®, a novel encapsulated sterile guidewire: A prospective cohort trial
Abstract
Purpose: We evaluated the performance of an encapsulated guidewire designed for single-handed use with ultrasound-guided vascular access (SonoStik) with Seldinger technique, as compared with conventional intravenous catheters placed under ultrasound guidance in healthy subjects.
Methods: This is a prospective cohort trial in healthy subjects in which each subject served as his/her own control by having a SonoStik ultrasound intravenous cannulation placed in one arm and a conventionally placed, standard ultrasound intravenous cannulation placed in the other arm. The basilic vein was used because it is a non-visible and non-palpable vein. Emergency department technicians with extensive experience in ultrasound-guided intravenous access performed the procedures. The first-attempt success rate of intravenous-guided intravenous by using the SonoStik was compared to the standard ultrasound intravenous cannulation in adult healthy subjects. The secondary outcomes including time of procedure, technicians' and subjects' satisfaction, and complications were compared in both arms of the study.
Results: A total of 24 volunteers with a mean age of 22.7 years were enrolled. Four emergency department technicians with extensive prior experience with ultrasound-guided intravenous access but with no prior experience using the SonoStik device performed the procedures. The first-attempt success was 83.3% with the use of SonoStik ultrasound intravenous cannulation compared to 95.8% with the standard ultrasound intravenous cannulation. There was a mean of 1.14 insertions per each successful placement in the SonoStik group compared to 1.04 insertions by using the standard catheters (mean differences = -0.1; 95% confidence interval = -0.6 to 0.4). There were no complications in either SonoStik or the standard ultrasound intravenous cannulation group. The mean time of insertion using SonoStik was slightly longer compared to standard ultrasound intravenous cannulation (143.3 vs 109.7 s).
Conclusion: This study demonstrated that emergency department technicians skilled in ultrasound-guided intravenous access could successfully place SonoStik 83.3% of the time in vessels that were unable to be palpated or visualized. Compared to standard ultrasound intravenous cannulation, the odds ratio of successful cannulation with SonoStik was 0.91 (95% confidence interval = 0.04-17.5). In all cases, the time required to successfully insert SonoStik was less than 4 min from tourniquet application to catheter advancement to hub, with a mean time of less than 2.5 min.
Keywords: Intravenous access; SonoStik®; ultrasound.
Similar articles
-
A randomized controlled trial comparing one-operator versus two-operator technique in ultrasound-guided basilic vein cannulation.J Emerg Med. 2008 Nov;35(4):431-5. doi: 10.1016/j.jemermed.2007.11.009. Epub 2008 Feb 21. J Emerg Med. 2008. PMID: 18291612 Clinical Trial.
-
Ultrasound-guided brachial and basilic vein cannulation in emergency department patients with difficult intravenous access.Ann Emerg Med. 1999 Dec;34(6):711-4. doi: 10.1016/s0196-0644(99)70095-8. Ann Emerg Med. 1999. PMID: 10577399
-
"Guidewire Intravenous Catheter Systems Do Not Improve First-Pass Success Rates for Peripheral Access When Placed By Army Combat Medics (68W) in a Pre-hospital Setting." A Prospective, Randomized Controlled Trial with Crossover Study Design.Mil Med. 2018 Nov 1;183(11-12):e730-e734. doi: 10.1093/milmed/usy110. Mil Med. 2018. PMID: 29800299 Clinical Trial.
-
Comparison of ultrasound guidance with palpation and direct visualisation for peripheral vein cannulation in adult patients: a systematic review and meta-analysis.Br J Anaesth. 2018 Aug;121(2):358-366. doi: 10.1016/j.bja.2018.04.047. Epub 2018 Jul 2. Br J Anaesth. 2018. PMID: 30032874
-
Ultrasound-guided peripheral intravenous canulation by emergency nurses: A systematic review and meta-analysis.Int Emerg Nurs. 2024 Mar;73:101422. doi: 10.1016/j.ienj.2024.101422. Epub 2024 Feb 23. Int Emerg Nurs. 2024. PMID: 38401479
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources