Smartwatch Algorithm for Automated Detection of Atrial Fibrillation
- PMID: 29535065
- DOI: 10.1016/j.jacc.2018.03.003
Smartwatch Algorithm for Automated Detection of Atrial Fibrillation
Abstract
Background: The Kardia Band (KB) is a novel technology that enables patients to record a rhythm strip using an Apple Watch (Apple, Cupertino, California). The band is paired with an app providing automated detection of atrial fibrillation (AF).
Objectives: The purpose of this study was to examine whether the KB could accurately differentiate sinus rhythm (SR) from AF compared with physician-interpreted 12-lead electrocardiograms (ECGs) and KB recordings.
Methods: Consecutive patients with AF presenting for cardioversion (CV) were enrolled. Patients underwent pre-CV ECG along with a KB recording. If CV was performed, a post-CV ECG was obtained along with a KB recording. The KB interpretations were compared to physician-reviewed ECGs. The KB recordings were reviewed by blinded electrophysiologists and compared to ECG interpretations. Sensitivity, specificity, and K coefficient were measured.
Results: A total of 100 patients were enrolled (age 68 ± 11 years). Eight patients did not undergo CV as they were found to be in SR. There were 169 simultaneous ECG and KB recordings. Fifty-seven were noninterpretable by the KB. Compared with ECG, the KB interpreted AF with 93% sensitivity, 84% specificity, and a K coefficient of 0.77. Physician interpretation of KB recordings demonstrated 99% sensitivity, 83% specificity, and a K coefficient of 0.83. Of the 57 noninterpretable KB recordings, interpreting electrophysiologists diagnosed AF with 100% sensitivity, 80% specificity, and a K coefficient of 0.74. Among 113 cases where KB and physician readings of the same recording were interpretable, agreement was excellent (K coefficient = 0.88).
Conclusions: The KB algorithm for AF detection supported by physician review can accurately differentiate AF from SR. This technology can help screen patients prior to elective CV and avoid unnecessary procedures.
Keywords: ECG monitoring; atrial fibrillation; cardioversion; digital health; smartwatch.
Copyright © 2018 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Comment in
-
Smartwatches in the Fight Against Atrial Fibrillation: The Little Watch That Could.J Am Coll Cardiol. 2018 May 29;71(21):2389-2391. doi: 10.1016/j.jacc.2018.03.485. J Am Coll Cardiol. 2018. PMID: 29793627 No abstract available.
-
The Smartwatch Will See You Now: Implications of Mass Screening for Atrial Fibrillation.J Am Coll Cardiol. 2018 Sep 18;72(12):1433-1434. doi: 10.1016/j.jacc.2018.05.078. J Am Coll Cardiol. 2018. PMID: 30213343 No abstract available.
-
Reply: The Smartwatch Will See You Now: Implications of Mass Screening for Atrial Fibrillation.J Am Coll Cardiol. 2018 Sep 18;72(12):1434-1435. doi: 10.1016/j.jacc.2018.07.022. J Am Coll Cardiol. 2018. PMID: 30213344 No abstract available.
-
Smart mobile devices in health care-smart enough to detect atrial fibrillation?J Thorac Dis. 2018 Sep;10(Suppl 26):S3227-S3228. doi: 10.21037/jtd.2018.08.21. J Thorac Dis. 2018. PMID: 30370121 Free PMC article. No abstract available.
-
Smart devices for a smart detection of atrial fibrillation.J Thorac Dis. 2018 Nov;10(Suppl 33):S3824-S3827. doi: 10.21037/jtd.2018.08.138. J Thorac Dis. 2018. PMID: 30631488 Free PMC article. No abstract available.
-
A case for the smart use of smartwatch-based technologies.J Thorac Dis. 2018 Nov;10(Suppl 33):S3875-S3877. doi: 10.21037/jtd.2018.08.141. J Thorac Dis. 2018. PMID: 30631503 Free PMC article. No abstract available.