Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Multicenter Study
. 2018 May;39(5):841-847.
doi: 10.3174/ajnr.A5592. Epub 2018 Mar 15.

European Multicenter Study for the Evaluation of a Dual-Layer Flow-Diverting Stent for Treatment of Wide-Neck Intracranial Aneurysms: The European Flow-Redirection Intraluminal Device Study

Affiliations
Multicenter Study

European Multicenter Study for the Evaluation of a Dual-Layer Flow-Diverting Stent for Treatment of Wide-Neck Intracranial Aneurysms: The European Flow-Redirection Intraluminal Device Study

M Killer-Oberpfalzer et al. AJNR Am J Neuroradiol. 2018 May.

Abstract

Background and purpose: Endoluminal reconstruction with flow-diverting stents represents a widely accepted technique for the treatment of complex intracranial aneurysms. This European registry study analyzed the initial experience of 15 neurovascular centers with the Flow-Redirection Intraluminal Device (FRED) system.

Materials and methods: Consecutive patients with intracranial aneurysms treated with the FRED between February 2012 and March 2015 were retrospectively reviewed. Complications and adverse events, transient and permanent morbidity, mortality, and occlusion rates were evaluated.

Results: During the defined study period, 579 aneurysms in 531 patients (median age, 54 years; range, 13-86 years) were treated with the FRED. Seven percent of patients were treated in the acute phase (≤3 days) of aneurysm rupture. The median aneurysm size was 7.6 mm (range, 1-36.6 mm), and the median neck size 4.5 mm (range, 1-30 mm). Angiographic follow-up of >3 months was available for 516 (89.1%) aneurysms. There was progressive occlusion witnessed with time, with complete occlusion in 18 (20%) aneurysms followed for up to 90 ± 14 days, 141 (82.5%) for 180 ± 20 days, 116 (91.3%) for 1 year ± 24 days, and 122 (95.3%) aneurysms followed for >1 year. Transient and permanent morbidity occurred in 3.2% and 0.8% of procedures, respectively. The overall mortality rate was 1.5%.

Conclusions: This retrospective study in real-world patients demonstrated the safety and efficacy of the FRED for the treatment of intracranial aneurysms. In most cases, treatment with a single FRED resulted in complete angiographic occlusion at 1 year.

PubMed Disclaimer

References

    1. Becske T, Kallmes DF, Saatci I, et al. . Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial. Radiology 2013;267:858–68 10.1148/radiol.13120099 - DOI - PubMed
    1. Briganti F, Leone G, Cirillo L, et al. . Postprocedural, midterm, and long-term results of cerebral aneurysms treated with flow-diverter devices: 7-year experience at a single center. Neurosurg Focus 2017;42:E3 10.3171/2017.3.FOCUS1732 - DOI - PubMed
    1. Pivotal Study of the FRED Stent System in the Treatment of Intracranial Aneurysms. ClinicalTrials.gov. https://www.clinicaltrials.gov/ct2/show/NCT01801007?term=FRED&cond=Aneur.... Accessed December 18, 2017
    1. Luecking H, Engelhorn T, Lang S, et al. . FRED flow diverter: a study on safety and efficacy in a consecutive group of 50 patients. AJNR Am J Neuroradiol 2017;38:596–602 10.3174/ajnr.A5052 - DOI - PMC - PubMed
    1. Möhlenbruch MA, Herweh C, Jestaedt L, et al. . The FRED flow-diverter stent for intracranial aneurysms: clinical study to assess safety and efficacy. AJNR Am J Neuroradiol 2015;36:1155–61 10.3174/ajnr.A4251 - DOI - PMC - PubMed

Publication types

LinkOut - more resources