. 2018 Apr;46(4):715-721.
doi: 10.1016/j.jcms.2018.02.010. Epub 2018 Feb 26.

The advantages of advanced computer-assisted diagnostics and three-dimensional preoperative planning on implant position in orbital reconstruction

Affiliations

The advantages of advanced computer-assisted diagnostics and three-dimensional preoperative planning on implant position in orbital reconstruction

Jesper Jansen et al. J Craniomaxillofac Surg. 2018 Apr.

Abstract

Objective: Advanced three-dimensional (3D) diagnostics and preoperative planning are the first steps in computer-assisted surgery (CAS). They are an integral part of the workflow, and allow the surgeon to adequately assess the fracture and to perform virtual surgery to find the optimal implant position. The goal of this study was to evaluate the accuracy and predictability of 3D diagnostics and preoperative virtual planning without intraoperative navigation in orbital reconstruction.

Methods: In 10 cadaveric heads, 19 complex orbital fractures were created. First, all fractures were reconstructed without preoperative planning (control group) and at a later stage the reconstructions were repeated with the help of preoperative planning. Preformed titanium mesh plates were used for the reconstructions by two experienced oral and maxillofacial surgeons. The preoperative virtual planning was easily accessible for the surgeon during the reconstruction. Computed tomographic scans were obtained before and after creation of the orbital fractures and postoperatively. Using a paired t-test, implant positioning accuracy (translation and rotations) of both groups were evaluated by comparing the planned implant position with the position of the implant on the postoperative scan.

Results: Implant position improved significantly (P < 0.05) for translation, yaw and roll in the group with preoperative planning (Table 1). Pitch did not improve significantly (P = 0.78).

Conclusion: The use of 3D diagnostics and preoperative planning without navigation in complex orbital wall fractures has a positive effect on implant position. This is due to a better assessment of the fracture, the possibility of virtual surgery and because the planning can be used as a virtual guide intraoperatively. The surgeon has more control in positioning the implant in relation to the rim and other bony landmarks.

Keywords: Computer-assisted surgery; Implant position; Orbital fractures; Orbital reconstruction; Preoperative planning.

PubMed Disclaimer

MeSH terms

LinkOut - more resources