Observational Study
. 2018 Nov;19(6):542-547.
doi: 10.1177/1129729818761292. Epub 2018 Mar 18.

The intracavitary electrocardiography method for positioning the tip of epicutaneous cava catheter in neonates: Pilot study

Affiliations
Observational Study

The intracavitary electrocardiography method for positioning the tip of epicutaneous cava catheter in neonates: Pilot study

Antonella Capasso et al. J Vasc Access. 2018 Nov.

Abstract

Purpose:: The neonatologists of Sant'Anna and San Sebastiano Hospital of Caserta have carried out a pilot study investigating the safety, feasibility, and accuracy of intracavitary electrocardiography for neonatal epicutaneous cava catheter tip positioning.

Patients and methods:: We enrolled 39 neonates (1-28 days of postnatal age or correct age lower than 41 weeks) requiring epicutaneous cava catheter in the district of superior vena cava (head-neck or upper limbs). Intracavitary electrocardiography was applicable in 38 neonates.

Results:: No significant complications related to intracavitary electrocardiography occurred in the studied neonates. The increase in P wave on intracavitary electrocardiography was detected in 30 cases. Of the remaining eight cases, six malpositioned catheters tipped out of cavoatrial junction-target zone (chest x-ray and echocardiographical control) and two were false negative (tip located in target zone). The match between intracavitary electrocardiography and x-ray was observed in 29/38 cases, and the same ratio between intracavitary electrocardiography and echocardiography was detected.

Conclusion:: We conclude that the intracavitary electrocardiography method is safe and accurate in neonates as demonstrated in pediatric and adult patients. The applicability of the method is 97% and its feasibility is 79%. The overall accuracy is 76% but it rises to 97% if "peak" P wave is detected.

Keywords: Epicutaneous cava catheter; intracavitary electrocardiography; neonates; peripherally inserted central catheter; target zone; tip position.