Continuous noninvasive arterial blood pressure monitoring using the vascular unloading technology during complex gastrointestinal endoscopy: a prospective observational study

Affiliations
Observational Study

Continuous noninvasive arterial blood pressure monitoring using the vascular unloading technology during complex gastrointestinal endoscopy: a prospective observational study

Julia Y Nicklas et al. J Clin Monit Comput. 2019 Feb.

Abstract

The innovative vascular unloading technology (VUT) allows continuous noninvasive arterial blood pressure (AP) monitoring. We aimed to investigate whether the VUT enables AP changes to be detected earlier compared with intermittent AP monitoring in patients undergoing gastrointestinal endoscopy. In this prospective observational study, we recorded continuous AP measurements with the VUT (CNAP system; CNSystems Medizintechnik AG, Graz, Austria) and intermittent AP measurements with upper arm cuff oscillometry in 90 patients undergoing complex gastrointestinal endoscopy (Department of Interventional Endoscopy at the University Medical Center Hamburg-Eppendorf, Hamburg, Germany). A "hypotensive phase" was defined as a time period of at least 30 s during which ≥ 50% of the VUT-AP values were in a predefined range of hypotension, i.e., AP value a) ≥ 10% below the last oscillometric value and b) ≤ 65 mmHg for mean AP or ≤ 90 mmHg for systolic AP. In the 5-min-interval between two oscillometric measurements, one or more hypotensive phases were detected in 26 patients (29%) for mean AP and in 27 patients (30%) for systolic AP. Hypotensive phases had a mean duration of 195 ± 99 s for mean AP and 197 ± 97 s for systolic AP with a mean procedure duration of 36 (± 21) min. Continuous noninvasive AP monitoring using the VUT enables hypotensive phases to be detected earlier compared with intermittent AP monitoring during complex gastrointestinal endoscopy. These hypotensive phases may be missed or only belatedly recognized with intermittent AP monitoring. Continuous noninvasive AP measurement facilitates detecting hemodynamic instability more rapidly and therefore may improve patient safety.

Keywords: Arterial pressure; Hemodynamic monitoring; Hypotension; Oscillometry; Volume clamp method.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Gastrointest Endosc. 2007 Jul;66(1):27-34 - PubMed
    1. Int J Med Sci. 2009;6(1):37-42 - PubMed
    1. Br J Anaesth. 2010 Sep;105(3):264-72 - PubMed
    1. Br J Anaesth. 2012 Feb;108(2):202-10 - PubMed
    1. Can J Gastroenterol. 2012 Feb;26(2):71-8 - PubMed

Publication types

LinkOut - more resources