Correlation of Noninvasive Blood Pressure and Invasive Intra-arterial Blood Pressure in Patients Treated with Vasoactive Medications in a Neurocritical Care Unit
- PMID: 29569130
- DOI: 10.1007/s12028-018-0521-0
Correlation of Noninvasive Blood Pressure and Invasive Intra-arterial Blood Pressure in Patients Treated with Vasoactive Medications in a Neurocritical Care Unit
Abstract
Background: The correlation between noninvasive (oscillometric) blood pressure (NBP) and intra-arterial blood pressure (IAP) in critically ill patients receiving vasoactive medications in a Neurocritical Care Unit has not been systematically studied. The purpose of this study is to examine the relationship between simultaneously measured NBP and IAP recordings in these patients.
Methods: Prospective observational study of patients (N = 70) admitted to a neurocritical care unit receiving continuous vasopressor or antihypertensive infusions. Paired NBP/IAP observations along with covariate and demographic data were abstracted via chart audit. Analysis was performed using SAS v9.4.
Results: A total of 2177 paired NBP/IAP observations from 70 subjects (49% male, 63% white, mean age 59 years) receiving vasopressors (n = 21) or antihypertensive agents (n = 49) were collected. Paired t test analysis showed significant differences between NBP versus IAP readings: ([systolic blood pressure (SBP): mean = 136 vs. 140 mmHg; p < 0.0001], [diastolic blood pressure (DBP): mean = 70 vs. 68 mmHg, p < 0.0001], [mean arterial blood pressure (MAP): mean = 86 vs. 90 mmHg, p < 0.0001]). Bland-Altman plots for MAP, SBP, and DBP demonstrate good inter-method agreement between paired measures (excluding outliers) and demonstrate NBP-IAP SBP differences at extremes of blood pressures. Pearson correlation coefficients show strong positive correlations for paired MAP (r = 0.82), SBP (r = 0.84), and DBP (r = 0.73) recordings. An absolute NBP-IAP SBP difference of > 20 mmHg was seen in ~ 20% of observations of nicardipine, ~ 25% of observations of norepinephrine, and ~ 35% of observations of phenylephrine. For MAP, the corresponding numbers were ~ 10, 15, and 25% for nicardipine, norepinephrine, and phenylephrine, respectively.
Conclusion: Despite overall strong positive correlations between paired NBP and IAP readings of MAP and SBP, clinically relevant differences in blood pressure are frequent. When treating with vasoactive infusions targeted to a specific BP goal, it is important to keep in mind that NBP and IAP values are not interchangeable.
Keywords: Blood pressure; Hemodynamics; Invasive arterial blood pressure monitoring; Monitoring; Neurocritical care; Pharmacology.
Similar articles
-
Intra-arterial versus oscillometric blood pressure measurements in women with severe peripartum hypertension undergoing urgent treatment with nicardipine: An observational study.Pregnancy Hypertens. 2021 Jun;24:100-106. doi: 10.1016/j.preghy.2021.03.003. Epub 2021 Mar 20. Pregnancy Hypertens. 2021. PMID: 33773326
-
[Comparison of invasive and non-invasive blood pressure in critically ill patients].Zhonghua Yi Xue Za Zhi. 2018 Oct 9;98(37):3005-3008. doi: 10.3760/cma.j.issn.0376-2491.2018.37.011. Zhonghua Yi Xue Za Zhi. 2018. PMID: 30392257 Chinese.
-
Comparison of noninvasive oscillometric and intra-arterial blood pressure measurements in hyperacute stroke.Blood Press Monit. 2007 Jun;12(3):149-56. doi: 10.1097/MBP.0b013e3280b083e2. Blood Press Monit. 2007. PMID: 17496464
-
[Problematic of intra-abdominal pressure measurement].Anaesthesist. 2009 May;58(5):527-31. doi: 10.1007/s00101-009-1533-0. Anaesthesist. 2009. PMID: 19384455 Review. German.
-
The critical care research networks experience.Neurocrit Care. 2012 Feb;16(1):20-8. doi: 10.1007/s12028-011-9610-z. Neurocrit Care. 2012. PMID: 21796493 Review.
Cited by
-
A comparison of invasive arterial blood pressure measurement with oscillometric non-invasive blood pressure measurement in patients with sepsis.J Anesth. 2024 Apr;38(2):222-231. doi: 10.1007/s00540-023-03304-2. Epub 2024 Feb 2. J Anesth. 2024. PMID: 38305914
-
Assessment of a simulation-based implementation method to support the introduction of a new invasive device: a prospective cross-over study.BMJ Open Qual. 2025 Jun 8;14(2):e003029. doi: 10.1136/bmjoq-2024-003029. BMJ Open Qual. 2025. PMID: 40484415 Free PMC article.
-
Agreement between non-invasive and invasive arterial blood pressure during surgery in the prone position: an error grid analysis.J Anesth. 2024 Dec;38(6):765-770. doi: 10.1007/s00540-024-03385-7. Epub 2024 Aug 13. J Anesth. 2024. PMID: 39138698
-
Will the Real Blood Pressure Please Stand Up?Neurocrit Care. 2018 Jun;28(3):263-264. doi: 10.1007/s12028-018-0555-3. Neurocrit Care. 2018. PMID: 29873024 No abstract available.
-
Discrepancy between invasive and non-invasive blood pressure readings in extremely preterm infants in the first four weeks of life.PLoS One. 2018 Dec 28;13(12):e0209831. doi: 10.1371/journal.pone.0209831. eCollection 2018. PLoS One. 2018. PMID: 30592742 Free PMC article.