. 2018 May:55:60-67.
doi: 10.1016/j.medengphy.2018.03.002. Epub 2018 Mar 22.

Simple and efficient thermal calibration for MEMS gyroscopes

Affiliations

Simple and efficient thermal calibration for MEMS gyroscopes

Alexis Nez et al. Med Eng Phys. 2018 May.

Abstract

Gyroscopes are now becoming one of the most sold MEMS sensors, given that the many applications that require their use are booming. In the medical field, gyroscopes can be found in Inertial Measurement Units used for the development of clinical tools that are dedicated to human-movement monitoring. However, MEMS gyroscopes are known to suffer from a drift phenomenon, which is mainly due to temperature variations. This drift dramatically affects measurement capability, especially that of cheap MEMs gyroscopes. Calibration is therefore a key factor in achieving accurate measurements. However, traditional calibration procedures are often complex and require costly equipment. This paper therefore proposes an easy protocol for performing a thermal gyroscope calibration. In this protocol, accuracy over the angular velocity is evaluated by referring to an optoelectronic measurement, and is compared with the traditional calibration performed by the manufacturer. The RMSE between the reference angular velocity and that obtained with the proposed calibration was of 0.7°/s, which was slightly smaller than the RMSE of 1.1°/s achieved by the manufacturer's calibration. An analysis of uncertainty propagation shows that offset variability is the major source of error over the computed rate of rotation from the tested sensors, since it accounts for 97% of the error. It can be concluded that the proposed simple calibration method leads to a similar degree of accuracy as that achieved by the manufacturer's procedure.

Keywords: Gyroscope; Measurement model; Microelectromechanical system (MEMS); Thermal calibration.