Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2018 May;20(5):344-352.
doi: 10.1089/dia.2018.0028. Epub 2018 Mar 30.

Interference Assessment of Various Endogenous and Exogenous Substances on the Performance of the Eversense Long-Term Implantable Continuous Glucose Monitoring System

Affiliations

Interference Assessment of Various Endogenous and Exogenous Substances on the Performance of the Eversense Long-Term Implantable Continuous Glucose Monitoring System

Carrie Lorenz et al. Diabetes Technol Ther. 2018 May.

Abstract

Background: A variety of prescriptions and over-the-counter medications interfere with transcutaneous continuous glucose monitoring (CGM) sensors. This study characterized the interference profile of the Eversense® CGM System (Senseonics, Inc., Germantown, MD), which has a different mechanism of glucose detection than other CGM systems.

Materials and methods: Sensor bias (sensor glucose concentration measurement - plasma glucose concentration measured by a reference test) was measured in vitro against 41 different substances at supratherapeutic/supraphysiologic plasma concentrations. Testing was performed using a paired-sample method adapted from the Clinical and Laboratory Standards Institute guidance document EP7-A2. Any substance producing sensor bias that exceeded the International Organization for Standardization (ISO) document 15197:2013 limits was then tested using an in vitro dose-response method to determine whether the concentration producing a significant sensor bias was within physiologic/therapeutic concentration ranges.

Results: Eight of 41 substances produced a sensor bias that exceeded ISO 15197:2013 limits when tested in vitro at supratherapeutic/supraphysiologic plasma concentrations. Only two of these substances (tetracycline and mannitol) exceeded bias limits within therapeutic concentration ranges. Notably, neither acetaminophen nor ascorbic acid, which are substances reported to interfere with other CGM systems, produced sensor bias that exceeded ISO limits when used at physiologic concentrations.

Conclusions: Although tetracycline and mannitol interfered with the Eversense sensor, substances frequently reported to interfere with enzymatic, electrochemical-based transcutaneous CGM systems, such as acetaminophen and ascorbic acid, did not affect Eversense readings.

Keywords: Continuous glucose monitoring; Drug interferences; Sensor.

PubMed Disclaimer

Conflict of interest statement

All the authors are employees of Senseonics, Inc., and have stock options or own stock in the company.

Figures

<b>FIG. 1.</b>
FIG. 1.
Eversense® implantable glucose sensor. DXA, dexamethasone acetate; PMMA, poly(methyl methacrylate).
<b>FIG. 2.</b>
FIG. 2.
Equilibrium binding of glucose to the glucose indicating polymer and effects on fluorescence. R2 denotes connectivity to the indicator polymer backbone. LED, light-emitting diode.

Similar articles

Cited by

References

    1. Parkin CG, Graham C, Smolskis J: Continuous glucose monitoring use in type 1 diabetes: longitudinal analysis demonstrates meaningful improvements in HbA1c and Reductions in health care utilization. J Diabetes Sci Technol 2017;11:522–528 - PMC - PubMed
    1. Battelino T, Conget I, Olsen B, et al. : The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial. Diabetologia 2012;55:3155–3162 - PMC - PubMed
    1. Beck RW, Riddlesworth T, Ruedy K, et al. : Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections: the DIAMOND randomized clinical trial. JAMA 2017;317:371–378 - PubMed
    1. Abraham MB, de Bock M, Paramalingam N, et al. : Prevention of insulin-induced hypoglycemia in type 1 diabetes with predictive low glucose management system. Diabetes Technol Ther 2016;18:436–443 - PubMed
    1. Pickup JC, Freeman SC, Sutton AJ: Glycaemic control in type 1 diabetes during real time continuous glucose monitoring compared with self monitoring of blood glucose: meta-analysis of randomised controlled trials using individual patient data. BMJ 2011;343:d3805. - PMC - PubMed

Publication types