. 1987 Sep-Oct;28(5):577-80.

Low dose pelvimetry with biplane digital radiography

Affiliations
  • PMID: 2960351

Low dose pelvimetry with biplane digital radiography

H Lotz et al. Acta Radiol. 1987 Sep-Oct.

Abstract

The digital moving slit technique as used by most CT scanners for overview images, was used for antero-posterior and lateral views of the pelvis for pelvimetry. The method was evaluated in phantom experiments and clinical examinations were performed in 23 patients. The method was compared with conventional pelvimetry in 14 patients. Estimated ovarian dose was reduced by a factor of 14.2. The discrepancy in measurements of the pelvic diameters by computed tomography compared with conventional pelvimetry was considered to be without practical importance from an obstetric point of view. Digital pelvimetry is easier for the patient, faster and the need for repeated exposures is eliminated. Based on these facts it is suggested that digital pelvimetry should replace the conventional method whenever possible.