. 2018 May;410(14):3361-3374.
doi: 10.1007/s00216-018-1031-9. Epub 2018 Apr 2.

Pipette-tip solid-phase extraction using polypyrrole as efficient adsorbent for extraction of avermectins and milbemycins in milk

Affiliations

Pipette-tip solid-phase extraction using polypyrrole as efficient adsorbent for extraction of avermectins and milbemycins in milk

Diego Hernando Ângulo Florez et al. Anal Bioanal Chem. 2018 May.

Abstract

In this work, we developed a HPLC method for the multidetermination of avermectins (AVM) (abamectin-ABA 1b and ABA 1a, eprinomectin-EPR, and ivermectin-IVM) and milbemycins (moxidectin-MOX) in milk samples using polypyrrole (PPy) as adsorbent material in pipette-tip solid-phase extraction (PT-PPy-SPE). PPy was characterized by scanning electron microscopy, thermogravimetric analysis, Fourier transform infrared spectroscopy, and X-ray diffraction and the data agreed with the literature. The sample preparation included the clean-up of the milk by protein precipitation (PP) with acetonitrile and extraction of the analytes by PT-PPy-SPE. The chromatographic method was developed in reverse phase and isocratic mode with flow rate at 1.2 mL min-1 and ultraviolet detection at 250 nm. The mobile phase composition was acetonitrile:methanol:water (55:25:20, v/v/v). The studied parameters and the optimized conditions for the sample preparation were washing solvent (300 μL water), volume and type of eluent (500 μL methanol), volume and pH of sample (1 mL and pH 10), amount of adsorbent material (50 mg PPy), and without addition of salt (NaCl). The method was linear over the concentration range from 20 to 3000 ng mL-1 with coefficients of correlation (r) ≥ 0.99 for all analytes and recoveries around 100%. The method developed and validated was used for the analyses of real milk samples from cow treated with Ivomec® (IVM 3.5%), in which were found 21.51 ± 2.94 ng mL-1 of IVM. Finally, the results proved that PT-PPy-SPE coupled to HPLC-UV was economical, simple, and easy-to-perform technique. Graphical abstract Pipette-tip solid phase extraction using polypirrole as adsorbent material for determination of avermectins and milbemycins in milk.

Keywords: Avermectins; Milbemycyns; Milk; Pipette-tip solid-phase extraction; Polypyrrole.