Experimental validation of robot-assisted cardiovascular catheterization: model-based versus model-free control

Affiliations

Experimental validation of robot-assisted cardiovascular catheterization: model-based versus model-free control

Xiaomei Wang et al. Int J Comput Assist Radiol Surg. 2018 Jun.

Abstract

Purpose: In cardiac electrophysiology, a long and flexible catheter is delivered to a cardiac chamber for the treatment of arrhythmias. Although several robot-assisted platforms have been commercialized, the disorientation in tele-operation is still not well solved. We propose a validation platform for robot-assisted cardiac EP catheterization, integrating a customized MR Safe robot, a standard clinically used EP catheter, and a human-robot interface. Both model-based and model-free control methods are implemented in the platform for quantitative evaluation and comparison.

Methods: The model-based and model-free control methods were validated by subject test (ten participants), in which the subjects have to perform a simulated radiofrequency ablation task using both methods. A virtual endoscopic view of the catheter is also provided to enhance hand-to-eye coordination. Assessment indices for targeting accuracy and efficiency were acquired for the evaluation.

Results: (1) Accuracy: The average distance measured from catheter tip to the closest lesion target during ablation of model-free method was 19.1% shorter than that of model-based control. (2) Efficiency: The model-free control reduced the total missed targets by 35.8% and the maximum continuously missed targets by 46.2%, both indices corresponded to a low p value ([Formula: see text]).

Conclusion: The model-free method performed better in terms of both accuracy and efficiency, indicating the model-free control could adapt to soft interaction with environment, as compared with the model-based control that does not consider contacts.

Keywords: Cardiac electrophysiology; Endoscopic view; MR Safe; Model-based control; Model-free control; Robotic catheterization.

PubMed Disclaimer

References

    1. IEEE Trans Biomed Eng. 2009 Mar;56(3):621-32 - PubMed
    1. Int J Comput Assist Radiol Surg. 2017 Jul;12 (7):1199-1207 - PubMed
    1. Catheter Cardiovasc Interv. 2006 Apr;67(4):637-43 - PubMed
    1. J Cardiovasc Electrophysiol. 2006 Sep;17 (9):951-6 - PubMed
    1. IEEE Int Conf Robot Autom. 2014;2014:4393-4398 - PubMed

Publication types

LinkOut - more resources