Effect of Software Version on the Accuracy of an Intraoral Scanning Device
Abstract
Purpose: To investigate the impact of software version on the accuracy of an intraoral scanning device.
Materials and methods: A master tooth was scanned with a high-precision optical scanner and then 10 times with a CEREC Omnicam scanner with software versions 4.4.0 and 4.4.4. Discrepancies were measured using quality control software.
Results: Mean deviation for 4.4.0 was 36.2 ± 35 μm and for 4.4.4 was 20.7 ± 14.2 μm (P ≤ .001).
Conclusion: Software version has a significant impact on the accuracy of an intraoral scanner. It is important that researchers also publish the software version of scanners when publishing their findings.