Allergic contact dermatitis caused by isobornyl acrylate in OmniPod, an innovative tubeless insulin pump

Affiliations
Case Reports

Allergic contact dermatitis caused by isobornyl acrylate in OmniPod, an innovative tubeless insulin pump

Nadia Raison-Peyron et al. Contact Dermatitis. 2018 Aug.

Abstract

Background: The frequency of allergic contact dermatitis caused by (meth)acrylates is increasing worldwide, and isobornyl acrylate (IBOA) has been identified as a culprit sensitizer in FreeStyle Libre, a medical device used by diabetic patients.

Objectives: To report on 4 patients sensitized to IBOA contained in OmniPod, a new, tubeless insulin pump, and to describe the clinical and chemical analytical work-up.

Method: Four patients with eczematous skin eruptions after using OmniPod, including 2 with previous adverse skin reactions to FreeStyle Libre, were patch tested with a baseline series and additional (meth)acrylates, including IBOA in several concentrations and/or vehicles. Gas chromatography-mass spectrometry (GC-MS) was used to identify the presence of IBOA in OmniPod.

Results: All patients were shown to be sensitized to IBOA. GC-MS identified IBOA in OmniPod, with the highest concentrations being found in the unit itself and not in the adhesive patch.

Conclusions: The OmniPod tubeless insulin pump is another source of IBOA, and its use may lead to primary sensitization to this acrylate, or may provoke elicitation of allergic contact dermatitis in previously sensitized patients. Our cases highlight the need to obtain satisfactory cooperation from pharmaceutical companies involved in the manufacture of these widely used medical devices.

Keywords: CAS no. 5888-33-5; OmniPod; adhesive; allergic contact dermatitis; case report; diabetes mellitus; gas chromatography-mass spectrometry; insulin pump; isobornyl acrylate; medical device.

PubMed Disclaimer

References

REFERENCES

    1. Bruze M, Trulsson L, Bendsoe N. Patch testing with ultrasonic batch extracts. Am J Contact Dermat. 1992;3:133-137.
    1. Herman A, Aerts O, Baeck M, et al. Allergic contact dermatitis caused by isobornyl acrylate in Freestyle® Libre, a newly introduced glucose sensor. Contact Dermatitis. 2017;77:367-373.
    1. Busschots AM, Meuleman V, Poesen N, Dooms-Goossens A. Contact allergy to components of glue in insulin pump infusion sets. Contact Dermatitis. 1995;33:205-206.
    1. Jolanki R, Kanerva L, Estlander T, et al. Allergic contact dermatitis from phenoxyethoxy ethylacrylates in optical fiber coating, and glue in an insulin pump set. Contact Dermatitis. 2001;45:36-37.
    1. Boom BW, van Driel LM. Allergic contact dermatitis to epoxy resin in infusion sets of an insulin pump. Contact Dermatitis. 1985;12:280.

Publication types

LinkOut - more resources