Abstract
Background: Safe and effective decontamination and reuse of N95 filtering facepiece respirators (FFRs) has the potential to significantly extend FFR holdings, mitigating a potential shortage due to an influenza pandemic or other pandemic events. Ultraviolet germicidal irradiation (UVGI) has been shown to be effective for decontaminating influenza-contaminated FFRs. This study aims to build on past research by evaluating the UVGI decontamination efficiency of influenza-contaminated FFRs in the presence of soiling agents using an optimized UVGI dose.
Methods: Twelve samples each of 15 N95 FFR models were contaminated with H1N1 influenza (facepiece and strap), then covered with a soiling agent-artificial saliva or artificial skin oil. For each soiling agent, 3 contaminated FFRs were treated with 1 J/cm2 UVGI for approximately 1 minute, whereas 3 other contaminated FFRs remained untreated. All contaminated surfaces were cut out and virus extracted. Viable influenza was quantified using a median tissue culture infectious dose assay.
Results: Significant reductions (≥3 log) in influenza viability for both soiling conditions were observed on facepieces from 12 of 15 FFR models and straps from 7 of 15 FFR models.
Conclusions: These data suggest that FFR decontamination and reuse using UVGI can be effective. Implementation of a UVGI method will require careful consideration of FFR model, material type, and design.
Keywords: Decontamination; Disinfection; Reuse; Soiling; UVGI; Ultraviolet.
Copyright © 2018 Association for Professionals in Infection Control and Epidemiology, Inc. All rights reserved.
Figures




Comment in
-
The importance of the minimum dosage necessary for UVC decontamination of N95 respirators during the COVID-19 pandemic.Photodermatol Photoimmunol Photomed. 2020 Jul;36(4):324-325. doi: 10.1111/phpp.12562. Epub 2020 Apr 28. Photodermatol Photoimmunol Photomed. 2020. PMID: 32291807 Free PMC article. No abstract available.
References
-
- Occupational Safety and Health Administration Pandemic Influenza Preparedness and Response Guidance for Healthcare Workers and Healthcare Employers. 2009. https://www.osha.gov/Publications/OSHA_pandemic_health.pdf OSHA 3328-05R; Available from.
-
- Centers for Disease Control Interim Domestic Guidance on the Use of Respirators to Prevent Transmission of SARS. https://www.cdc.gov/sars/clinical/respirators.html Available from.
-
- National Institute for Occupational Safety and Health NIOSH Guide to the Selection and Use of Particulate Respirators. https://www.cdc.gov/niosh/docs/96-101/default.html Available from.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous