Particle Accumulation in Ureteral Stents Is Governed by Fluid Dynamics: In Vitro Study Using a "Stent-on-Chip" Model
- PMID: 29699424
- DOI: 10.1089/end.2017.0946
Particle Accumulation in Ureteral Stents Is Governed by Fluid Dynamics: In Vitro Study Using a "Stent-on-Chip" Model
Abstract
Objective: To investigate the correlation between fluid dynamic processes and deposition of encrusting particles in ureteral stents.
Materials and methods: Microfluidic models (referred to as "stent-on-chip" or SOC) were developed to replicate relevant hydrodynamic regions of a stented ureter, including drainage holes and the cavity formed by a ureteral obstruction. Computational fluid dynamic simulations were performed to determine the wall shear stress (WSS) field over the solid surfaces of the model, and the computational flow field was validated experimentally. Artificial urine was conveyed through the SOCs to measure the temporal evolution of encrustation through optical microscopy.
Results: It was revealed that drainage holes located well downstream of the obstruction had almost stagnant flow and low WSS (average 0.01 Pa, at 1 mL/min), and thus suffered from higher encrustation rates. On the contrary, higher levels of WSS in holes proximal to the obstruction (average ∼0.04 Pa, at 1 mL/min) resulted in lower encrustation rates in these regions. The cavity located nearby the obstruction was characterized by high levels of encrustation, because of the low WSS (average 1.6 × 10-4 Pa, at 1 mL/min) and the presence of flow vortices. Increasing the drainage flow rate from 1 to 10 mL/min resulted in significantly lower deposition of encrusting crystals.
Conclusion: This study demonstrated an inverse correlation between deposition of encrusting bodies and the local WSS in a stented ureter model. Critical regions with low WSS and susceptible to encrustation were identified, including "inactive" side holes (i.e., with minimal or absent flow exchange between stent and ureter) and the cavity formed by a ureteral occlusion. Findings from this study can open new avenues for improving the stent's design through fluid dynamic optimization.
Keywords: encrustation; flow dynamics; microfluidic; side holes; ureter; ureteral stent.
Similar articles
-
Reducing deposition of encrustation in ureteric stents by changing the stent architecture: A microfluidic-based investigation.Biomicrofluidics. 2019 Jan 4;13(1):014101. doi: 10.1063/1.5059370. eCollection 2019 Jan. Biomicrofluidics. 2019. PMID: 30867872 Free PMC article.
-
Investigating the encrustation of reinforced ureteral stents by computational flow dynamic simulations.World J Urol. 2023 May;41(5):1451-1457. doi: 10.1007/s00345-023-04356-5. Epub 2023 Mar 17. World J Urol. 2023. PMID: 36930252 Free PMC article.
-
The accumulation of particles in ureteric stents is mediated by flow dynamics: Full-scale computational and experimental modeling of the occluded and unoccluded ureter.APL Bioeng. 2022 May 5;6(2):026102. doi: 10.1063/5.0083260. eCollection 2022 Jun. APL Bioeng. 2022. PMID: 35540726 Free PMC article.
-
Advances in Ureteral Stent Design and Materials.Curr Urol Rep. 2018 Apr 10;19(5):35. doi: 10.1007/s11934-018-0779-y. Curr Urol Rep. 2018. PMID: 29637309 Free PMC article. Review.
-
Ureteral Stent Encrustation: Epidemiology, Pathophysiology, Management and Current Technology.J Urol. 2021 Jan;205(1):68-77. doi: 10.1097/JU.0000000000001343. Epub 2020 Aug 28. J Urol. 2021. PMID: 32856981 Review.
Cited by
-
Reducing deposition of encrustation in ureteric stents by changing the stent architecture: A microfluidic-based investigation.Biomicrofluidics. 2019 Jan 4;13(1):014101. doi: 10.1063/1.5059370. eCollection 2019 Jan. Biomicrofluidics. 2019. PMID: 30867872 Free PMC article.
-
The interplay between bacterial biofilms, encrustation, and wall shear stress in ureteral stents: a review across scales.Front Urol. 2024 Jan 16;3:1335414. doi: 10.3389/fruro.2023.1335414. eCollection 2023. Front Urol. 2024. PMID: 40778079 Free PMC article. Review.
-
Strategies to Improve Patient Outcomes and QOL: Current Complications of the Design and Placements of Ureteric Stents.Res Rep Urol. 2020 Jul 31;12:303-314. doi: 10.2147/RRU.S233981. eCollection 2020. Res Rep Urol. 2020. PMID: 32802807 Free PMC article. Review.
-
Application of microfluidic chips in the simulation of the urinary system microenvironment.Mater Today Bio. 2023 Jan 20;19:100553. doi: 10.1016/j.mtbio.2023.100553. eCollection 2023 Apr. Mater Today Bio. 2023. PMID: 36747584 Free PMC article. Review.
-
Computational simulation of the flow dynamic field in a porous ureteric stent.Med Biol Eng Comput. 2022 Aug;60(8):2373-2387. doi: 10.1007/s11517-022-02620-1. Epub 2022 Jun 28. Med Biol Eng Comput. 2022. PMID: 35763188 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials