The Doppler signal: where does it come from and what does it mean?
- PMID: 2970215
- DOI: 10.2214/ajr.151.3.439
The Doppler signal: where does it come from and what does it mean?
Abstract
Doppler sonographic measurement of blood velocity and associated physiologic parameters is a powerful diagnostic technique. State-of-the-art instrumentation incorporates velocity measurement with two-dimensional imaging capability; it uses intensity and color coding to display complex physiologic and anatomic data to the observer in an easily understood format. Although the concepts underlying Doppler sonography are not complex, mastery of the technique requires extra training and commitment. The principal features and clinical practicalities associated with Doppler sonography are summarized in the following paragraphs. Continuous-wave Doppler is very sensitive to small vessels and has no upper velocity limit. In addition, the instrumentation is not complex and produces relatively low acoustic power. A significant drawback to continuous-wave Doppler is that there is no depth sensitivity, and thus complex structures or multiple vessels can give conflicting information. Pulsed Doppler (including duplex and color-flow) instrumentation has the capability of depth resolution and a variable sample volume. Pulsed Doppler equipment is prone to aliasing (false velocity indications) under some circumstances and also produces higher peak power levels than does continuous-wave equipment. Duplex equipment is more complex and expensive than continuous-wave equipment because the two-dimensional and Doppler modes must be synchronized in operation and display. Color-flow equipment is extremely complex and expensive. Color flow provides information of a qualitative and limited quantitative value. Absolute measurement still requires range-gate measurements. Technical and anatomic factors will affect the measured velocity profiles. Thus, it is important to fully appreciate the anatomy of the vessel and the angle between the vessel and the ultrasound beam when making quantitative measurements. Measurements that evaluate the velocity waveform and make use of ratios, such as the pulsatility index, eliminate the need for angular corrections; however, artifacts due to unappreciated anatomic or wall characteristics may lead to incorrect information if all parameters are not fully understood. Doppler sonographic measurements may be used to determine the presence of flow, determine the direction of flow, identify time-varying velocity characteristics, and detect velocity disturbances. Because flow and velocity are related, it is possible to estimate flow from velocity measurements with careful calibration and proper precautions. Velocity is related to flow, which, in turn, is related to both pressure and vascular resistance.(ABSTRACT TRUNCATED AT 400 WORDS)
Similar articles
-
Validation of a sensitivity performance index test protocol and evaluation of colour Doppler sensitivity for a range of ultrasound scanners.Ultrasound Med Biol. 2004 Nov;30(11):1475-83. doi: 10.1016/j.ultrasmedbio.2004.09.005. Ultrasound Med Biol. 2004. PMID: 15588958
-
Principles of Doppler and color flow.Radiol Med. 1993 May;85(5 Suppl 1):3-16. Radiol Med. 1993. PMID: 8332812 Review.
-
Recent advances in carotid artery evaluation.Clin Diagn Ultrasound. 1990;26:25-44. Clin Diagn Ultrasound. 1990. PMID: 2182090 Review.
-
Studies of blood flow employing duplex, pulsed Doppler real-time sonography.Eur J Radiol. 1983 Aug;3 Suppl 1:268-72. Eur J Radiol. 1983. PMID: 6628413
-
Doppler color imaging. Principles and instrumentation.Clin Diagn Ultrasound. 1992;27:7-60. Clin Diagn Ultrasound. 1992. PMID: 1497942 Review.
Cited by
-
Evaluating feline lower urinary tract disease: Doppler ultrasound of the kidneys.J Feline Med Surg. 2023 Jan;25(1):1098612X221145477. doi: 10.1177/1098612X221145477. J Feline Med Surg. 2023. PMID: 36649073 Free PMC article.
-
Spectral Doppler ultrasound in the major arteries of normal conscious immature micropigs.J Vet Sci. 2010 Jun;11(2):155-9. doi: 10.4142/jvs.2010.11.2.155. J Vet Sci. 2010. PMID: 20458157 Free PMC article.
-
A theoretical foundation for relating the velocity time integrals of the left ventricular outflow tract and common carotid artery.J Clin Monit Comput. 2023 Jun;37(3):937-939. doi: 10.1007/s10877-022-00969-0. Epub 2023 Jan 10. J Clin Monit Comput. 2023. PMID: 36625983 Free PMC article. No abstract available.
-
Doppler ultrasonography of the pectinis oculi artery in harpy eagles (Harpia harpyja).Open Vet J. 2017;7(1):70-74. doi: 10.4314/ovj.v7i1.11. Epub 2017 Mar 31. Open Vet J. 2017. PMID: 28540254 Free PMC article.
-
Vascular hemodynamic effects of penile revascularization surgery and the role of resistive index in follow-up.Basic Clin Androl. 2024 Dec 20;34(1):28. doi: 10.1186/s12610-024-00243-0. Basic Clin Androl. 2024. PMID: 39701967 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical