Localization of ventricular activation origin using patient-specific geometry: Preliminary results
- PMID: 29702740
- DOI: 10.1111/jce.13622
Localization of ventricular activation origin using patient-specific geometry: Preliminary results
Abstract
Background and objectives: Catheter ablation of ventricular tachycardia (VT) may include induction of VT and localization of VT-exit site. Our aim was to assess localization performance of a novel statistical pace-mapping method and compare it with performance of an electrocardiographic inverse solution.
Methods: Seven patients undergoing ablation of VT (4 with epicardial, 3 with endocardial exit) aided by electroanatomic mapping underwent intraprocedural 120-lead body-surface potential mapping (BSPM). Two approaches to localization of activation origin were tested: (1) A statistical method, based on multiple linear regression (MLR), which required only the conventional 12-lead ECG for a sufficient number of pacing sites with known origin together with patient-specific geometry of the endocardial/epicardial surface obtained by electroanatomic mapping; and (2) a classical deterministic inverse solution for recovering heart-surface potentials, which required BSPM and patient-specific geometry of the heart and torso obtained via computed tomography (CT).
Results: For the MLR method, at least 10-15 pacing sites with known coordinates, together with their corresponding 12-lead ECGs, were required to derive reliable patient-specific regression equations, which then enabled accurate localization of ventricular activation with unknown origin. For 4 patients who underwent epicardial mapping, the median of localization error for the MLR was significantly lower than that for the inverse solution (10.6 vs. 27.3 mm, P = 0.034); a similar result held for 3 patients who underwent endocardial mapping (7.7 vs. 17.1 mm, P = 0.017). The pooled localization error for all epicardial and endocardial sites was also significantly smaller for the MLR compared with the inverse solution (P = 0.005).
Conclusions: The novel pace-mapping approach to localizing the origin of ventricular activation offers an easily implementable supplement and/or alternative to the preprocedure inverse solution; its simplicity makes it suitable for real-time applications during clinical catheter-ablation procedures.
Keywords: 12-lead ECG; body surface potential mapping; catheter ablation; pace-mapping; ventricular tachycardia.
© 2018 Wiley Periodicals, Inc.
Comment in
-
Statistical guidance of VT ablation.J Cardiovasc Electrophysiol. 2018 Jul;29(7):987-989. doi: 10.1111/jce.13633. Epub 2018 Jun 7. J Cardiovasc Electrophysiol. 2018. PMID: 29771455 Free PMC article. No abstract available.
Publication types
MeSH terms
Grants and funding