A high rate of tibial plateau fractures after early experience with patient-specific instrumentation for unicompartmental knee arthroplasties
- PMID: 29713783
- DOI: 10.1007/s00167-018-4956-9
A high rate of tibial plateau fractures after early experience with patient-specific instrumentation for unicompartmental knee arthroplasties
Abstract
Purpose: Patient-specific instrumentation (PSI) for unicompartmental knee arthroplasty (UKA) has been available for a few years. However, limited literature is available on this subject. Hence, the aim of this cohort study is to evaluate the 2 years' results of our first experiences with the use of PSI in UKA. It is hypothesised that there is no advantage in rate of adverse events and in radiological and functional outcomes in comparison to literature on the conventional method.
Methods: This cohort included 129 knees of 122 patients, operated by one surgeon. Outcome measures were the rate of adverse events (AEs); implant position as determined on radiographs; the accuracy of the default and approved planning of the implant sizes and the patient-reported outcome measures (PROMs) preoperatively, and at 3, 12 and 24 months, postoperatively.
Results: A total of 6 (4.9%) AEs were observed in this study, with 4 (3.3%) tibial fractures being the main complication. The mean postoperative biomechanical axis was 176.4° and in the majority of cases, the radiographic criteria, as determined by the manufacturer, were met. The tibial component showed 20 (16.4%) outliers in the sagittal and 3 (2.5%) outliers in the frontal plane. There were no outliers of the femoral component. For the femoral and tibial components, respectively, in 125 (96.9%) and 79 (61.7%) cases, there was an agreement between approved planning and implanted component size. All PROMs improved significantly after surgery.
Conclusion: Tibial fracture was the most common AE, probably related to the transition from cemented to uncemented UKA. Perioperative modifications to the surgical technique were made in order to prevent this AE. Improvements should be made to the operation technique of the uncemented tibial plateau to obtain an adequate placement and at the same time reduce the risk for tibial fracture. The PSI technique was a reliable tool for the placement of the femoral component. Functional outcome was in line with literature on the conventional method. It is strongly recommended that the surgeon approves every preoperative plan, in order to optimise the accuracy during the PSI surgery.
Level of evidence: III.
Keywords: Custom fit; Instruments; PSG; PSI; Patient specific; Patient-specific guides; Patient-specific matched instruments; Templating; UKA; Unicompartmental knee arthroplasty; Unicondylar knee arthroplasty.
Similar articles
-
Patient-specific instrumentation in Oxford unicompartmental knee arthroplasty is reliable and accurate except for the tibial rotation.Knee Surg Sports Traumatol Arthrosc. 2018 Jun;26(6):1823-1830. doi: 10.1007/s00167-017-4826-x. Epub 2017 Dec 27. Knee Surg Sports Traumatol Arthrosc. 2018. PMID: 29282486
-
Revision of partial knee to total knee arthroplasty with use of patient-specific instruments results in acceptable femoral rotation.Knee Surg Sports Traumatol Arthrosc. 2018 Jun;26(6):1656-1661. doi: 10.1007/s00167-017-4674-8. Epub 2017 Aug 7. Knee Surg Sports Traumatol Arthrosc. 2018. PMID: 28785788
-
The John Insall Award: No Functional Benefit After Unicompartmental Knee Arthroplasty Performed With Patient-specific Instrumentation: A Randomized Trial.Clin Orthop Relat Res. 2016 Jan;474(1):60-8. doi: 10.1007/s11999-015-4259-0. Clin Orthop Relat Res. 2016. PMID: 25791446 Free PMC article. Clinical Trial.
-
Patient-specific instrumentation does not improve radiographic alignment or clinical outcomes after total knee arthroplasty.Acta Orthop. 2016 Aug;87(4):386-94. doi: 10.1080/17453674.2016.1193799. Epub 2016 Jun 1. Acta Orthop. 2016. PMID: 27249110 Free PMC article. Review.
-
Patient-specific instrument for unicompartmental knee arthroplasty does not reduce the outliers in alignment or improve postoperative function: a meta-analysis and systematic review.Arch Orthop Trauma Surg. 2020 Aug;140(8):1097-1107. doi: 10.1007/s00402-020-03429-z. Epub 2020 Apr 18. Arch Orthop Trauma Surg. 2020. PMID: 32306092
Cited by
-
What future in the treatment of osteochondral knee defects?Ann Transl Med. 2018 Dec;6(Suppl 2):S100. doi: 10.21037/atm.2018.11.28. Ann Transl Med. 2018. PMID: 30740421 Free PMC article. No abstract available.
-
[New technologies (robotics, custom-made) in unicondylar knee arthroplasty-pro].Orthopade. 2021 Feb;50(2):130-135. doi: 10.1007/s00132-020-04058-5. Orthopade. 2021. PMID: 33346868 Review. German.
-
Association between bone marrow lesions and bone mineral density of the proximal tibia in end-stage osteoarthritic knees.Sci Rep. 2023 Apr 21;13(1):6516. doi: 10.1038/s41598-023-33251-7. Sci Rep. 2023. PMID: 37085519 Free PMC article.
-
Patient-specific instrumentation does not improve tibial component coronal alignment for medial UKA compared to conventional instrumentation.J Exp Orthop. 2020 Jun 8;7(1):42. doi: 10.1186/s40634-020-00257-3. J Exp Orthop. 2020. PMID: 32514889 Free PMC article.
-
Comparable incidence of periprosthetic tibial fractures in cementless and cemented unicompartmental knee arthroplasty: a systematic review and meta-analysis.Knee Surg Sports Traumatol Arthrosc. 2022 Mar;30(3):852-874. doi: 10.1007/s00167-021-06449-3. Epub 2021 Feb 2. Knee Surg Sports Traumatol Arthrosc. 2022. PMID: 33528591 Free PMC article.