MX2: a high-flux undulator microfocus beamline serving both the chemical and macromolecular crystallography communities at the Australian Synchrotron
- PMID: 29714201
- PMCID: PMC5929359
- DOI: 10.1107/S1600577518003120
MX2: a high-flux undulator microfocus beamline serving both the chemical and macromolecular crystallography communities at the Australian Synchrotron
Abstract
MX2 is an in-vacuum undulator-based crystallography beamline at the 3 GeV Australian Synchrotron. The beamline delivers hard X-rays in the energy range 4.8-21 keV to a focal spot of 22 × 12 µm FWHM (H × V). At 13 keV the flux at the sample is 3.4 × 1012 photons s-1. The beamline endstation allows robotic handling of cryogenic samples via an updated SSRL SAM robot. This beamline is ideal for weakly diffracting hard-to-crystallize proteins, virus particles, protein assemblies and nucleic acids as well as smaller molecules such as inorganic catalysts and organic drug molecules. The beamline is now mature and has enjoyed a full user program for the last nine years. This paper describes the beamline status, plans for its future and some recent scientific highlights.
Keywords: EIGER detector; anomalous scattering; apertures; collimators; long wavelengths; macromolecular crystallography; microfocus beamlines; remote access; undulators.
open access.
Figures
References
-
- Birch, W. D., Grey, I. E., Mills, S. J., Pring, A., Bougerol, C., Ribaldi-Tunnicliffe, A., Wilson, N. C. & Keck, E. (2011). Mineral. Mag. 75, 269–278.
-
- Boldeman, J. W. & Einfeld, D. (2004). Nucl. Instrum. Methods Phys. Res. A, 521, 306–317.
-
- Cohen, A. E., McPhillips, S. E., Song, J. & Miller, M. D. (2005). Synchrotron Rad. News, 18, 28–35.