. 2018 Feb;4(2):201-208.
doi: 10.1016/j.jacep.2017.09.178. Epub 2017 Nov 15.

Impact of Abandoned Leads on Cardiovascular Implantable Electronic Device Infections: A Propensity Matched Analysis of MEDIC (Multicenter Electrophysiologic Device Infection Cohort)

Affiliations
Free article

Impact of Abandoned Leads on Cardiovascular Implantable Electronic Device Infections: A Propensity Matched Analysis of MEDIC (Multicenter Electrophysiologic Device Infection Cohort)

Thomas A Boyle et al. JACC Clin Electrophysiol. 2018 Feb.
Free article

Abstract

Objectives: This study sought to evaluate the impact of abandoned cardiovascular implantable electronic device (CIED) leads on the presentation and management of device-related infections.

Background: Device infection is a serious consequence of CIEDs and necessitates removal of all hardware for attempted cure. The merits of extracting or retaining presumed sterile but nonfunctioning leads is a subject of ongoing debate.

Methods: The MEDIC (Multicenter Electrophysiologic Device Infection Cohort) prospectively enrolled patients with CIED infections at 10 institutions in the United States and abroad between January 1, 2009, and December 31, 2012. Within a propensity-matched cohort, relevant clinical information was compared between patients who had 1 or more abandoned leads at the time of infection and those who had none.

Results: Matching produced a cohort of 264 patients, including 176 with no abandoned leads and 88 with abandoned leads. The groups were balanced with respect to Charlson comorbidity index, oldest lead age, device type, sex, and race. At the time of admission, those with abandoned leads were less likely to demonstrate systemic signs of infection, including leukocytosis (p = 0.023) and positive blood cultures (p = 0.005). Conversely, patients with abandoned leads were more likely to demonstrate local signs of infections, including skin erosion (p = 0.031) and positive pocket cultures (p = 0.015). In addition, patients with abandoned leads were more likely to require laser extraction (p = 0.010).

Conclusions: The results of a large prospective registry of CIED infections demonstrated that patients with abandoned leads may present with different signs, symptoms, and microbiological findings and require laser extraction more than those without abandoned leads.

Keywords: abandoned leads; endocarditis; lead capping; lead extraction.

PubMed Disclaimer

Comment in

  • An Unpleasant Legacy: Infected Abandoned Leads.
    Lau CP, Tse HF. Lau CP, et al. JACC Clin Electrophysiol. 2018 Feb;4(2):209-211. doi: 10.1016/j.jacep.2017.10.007. Epub 2017 Nov 29. JACC Clin Electrophysiol. 2018. PMID: 29749939 No abstract available.