Effect of increasing flow when grooving during phacoemulsification
- PMID: 29773457
- DOI: 10.1016/j.jcrs.2018.01.034
Effect of increasing flow when grooving during phacoemulsification
Abstract
Purpose: To determine optimum flow settings on the Centurion Vision System during the grooving step in cataract surgery.
Setting: John A. Moran Eye Center Laboratory, University of Utah, Salt Lake City, Utah, USA.
Design: Experimental study.
Methods: Intact porcine lenses hardened by formalin and placed in a chamber designed to simulate the anterior chamber of the eye were used to test flow rate settings at 20 mL/min, 40 mL/min, and 60 mL/min. Vacuum was set at 400 mm Hg, longitudinal power at 80%, torsional power at 80%, and intraocular pressure at 50 mm Hg. A balanced phaco tip with a 20-degree tip and a 30-degree bevel was used. Efficiency (time to groove the lens in half) was determined.
Results: Increasing flow from 20 to 40 mL/min during grooving increased efficiency by 17% (P = .05), with no significant improvement shown at 60 mL/min.
Conclusions: A flow rate of 40 mL/min was determined to be most efficient during the grooving step of cataract surgery. Further increases in flow rate showed no statistically significant improvement in efficiency, and with only 17% improvement flow rates less than 40 mL/min might be almost as efficient and might be safer.
Copyright © 2018 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.
Similar articles
-
The effect of increasing power when grooving using phacoemulsification.Clin Ophthalmol. 2019 Apr 12;13:611-615. doi: 10.2147/OPTH.S194731. eCollection 2019. Clin Ophthalmol. 2019. PMID: 31043766 Free PMC article.
-
High vacuum and aspiration on phacoemulsification efficiency and chatter for Centurion.Can J Ophthalmol. 2019 Feb;54(1):136-138. doi: 10.1016/j.jcjo.2018.03.009. Epub 2018 May 18. Can J Ophthalmol. 2019. PMID: 30851768
-
Comparison of Vacuum and Aspiration on Phacoemulsification Efficiency and Chatter Using a Monitored Forced Infusion System.Am J Ophthalmol. 2016 Sep;169:162-167. doi: 10.1016/j.ajo.2016.06.030. Epub 2016 Jun 25. Am J Ophthalmol. 2016. PMID: 27349412
-
Determining optimal torsional ultrasound power for cataract surgery with automatic longitudinal pulses at maximum vacuum ex vivo.Am J Ophthalmol. 2014 Dec;158(6):1262-1266.e2. doi: 10.1016/j.ajo.2014.08.034. Epub 2014 Aug 28. Am J Ophthalmol. 2014. PMID: 25174892
-
Phacoemulsification in review: Optimization of cataract removal in an in vitro setting.Surv Ophthalmol. 2019 Nov-Dec;64(6):868-875. doi: 10.1016/j.survophthal.2019.06.007. Epub 2019 Jul 2. Surv Ophthalmol. 2019. PMID: 31276738 Review.
Cited by
-
Comment on: Experimental study comparing 2 different phacoemulsification systems with intraocular pressure control during steady-state flow and occlusion break surge events.J Cataract Refract Surg. 2024 Mar 1;50(3):312. doi: 10.1097/j.jcrs.0000000000001391. J Cataract Refract Surg. 2024. PMID: 38192058 Free PMC article. No abstract available.
-
Effect of Modified Peristaltic Phacoemulsification Device on Efficiency and Post-Occlusion Pressure Surge.Clin Ophthalmol. 2023 Aug 14;17:2341-2347. doi: 10.2147/OPTH.S416604. eCollection 2023. Clin Ophthalmol. 2023. PMID: 37600145 Free PMC article.
-
Application of the Active-Fluidics System in Phacoemulsification: A Review.J Clin Med. 2023 Jan 12;12(2):611. doi: 10.3390/jcm12020611. J Clin Med. 2023. PMID: 36675540 Free PMC article. Review.
-
The effect of increasing power when grooving using phacoemulsification.Clin Ophthalmol. 2019 Apr 12;13:611-615. doi: 10.2147/OPTH.S194731. eCollection 2019. Clin Ophthalmol. 2019. PMID: 31043766 Free PMC article.
-
Optimization of Phacoemulsification Tip Gauge on the Oertli CataRhex3 in an in vitro Setting.Clin Ophthalmol. 2022 Apr 8;16:1091-1097. doi: 10.2147/OPTH.S355094. eCollection 2022. Clin Ophthalmol. 2022. PMID: 35422610 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources