Randomized Controlled Trial
. 2018 Aug;11(8):1109-1118.
doi: 10.1016/j.jcmg.2018.03.003. Epub 2018 May 16.

Single Versus Standard Multiview Assessment of Global Longitudinal Strain for the Diagnosis of Cardiotoxicity During Cancer Therapy

Collaborators, Affiliations
Free article
Randomized Controlled Trial

Single Versus Standard Multiview Assessment of Global Longitudinal Strain for the Diagnosis of Cardiotoxicity During Cancer Therapy

Paaladinesh Thavendiranathan et al. JACC Cardiovasc Imaging. 2018 Aug.
Free article

Abstract

Objectives: The goal of this study was to compare echocardiographic measurements of global longitudinal strain (GLS) (using 3 apical views) with single-view longitudinal strain (LS, 4- or 2-chamber [4CV_LS and 2CV_LS, respectively]) for detection of cancer-therapy related cardiotoxicity.

Background: GLS is useful for the detection of cardiotoxicity, but the need for repeated measurements poses a significant burden on busy echocardiography laboratories. A single-view LS measurement, possibly at point of care, could improve efficiency.

Methods: Seventeen international centers prospectively recruited 108 patients (mean age 54 ± 13 years) at high risk for cardiotoxicity as part of the ongoing SUCCOUR (Strain Surveillance for Improving Cardiovascular Outcomes During Chemotherapy) randomized controlled trial. Echocardiography performed at baseline and follow-up were analyzed in a core laboratory setting blinded to clinical information. Peak systolic GLS and LS were measured from raw data. Cardiotoxicity was defined by reduction in left ventricular ejection fraction >0.10 to <0.55 or a relative drop in GLS by ≥12%.

Results: Cardiotoxicity developed in 46 patients by either criteria. Baseline and follow-up 2-dimensional left ventricular ejection fraction were 61 ± 4% and 58 ± 5%, respectively (p < 0.001). The baseline GLS (-20.9 ± 2.4%) was not different from 4CV_LS (-20.7 ± 2.5%; p = 0.09) or 2CV_LS (-21.1 ± 3.1%; p = 0.25). The follow-up GLS (-19.5 ± 2.4%) was also similar to 4CV_LS (-19.5 ± 2.6%; p = 0.80) and 2CV_LS (-19.7 ± 3.1%; p = 0.19). There was good correlation between GLS and 4CV_LS at baseline (r = 0.86; p < 0.001) and follow-up (r = 0.89; p < 0.001) and with 2CV_LS at baseline (r = 0.87; p < 0.001) and follow-up (r = 0.88; p < 0.001). However, there was 15% to 22% disagreement between GLS and 4CV_LS or 2CV_LS for the detection of cardiotoxicity. The interobserver and intraobserver reproducibility was higher for GLS (intraclass correlation: 0.93 to 0.95; coefficient of variance: 2.9% to 3.7%) compared with either single-chamber-based LS measurement (intraclass correlation: 0.85 to 0.91; coefficient of variance: 4.1% to 4.8%).

Conclusions: Although there was good correlation between GLS and single-view LS measurements, single-view LS measurement led to disagreement in the diagnosis of cardiotoxicity in up to 22% of patients. GLS measurements were more reproducible than single-view LS. GLS based on 3 apical views should remain the preferred technique for detection of cardiotoxicity. (Strain Surveillance for Improving Cardiovascular Outcomes During Chemotherapy [SUCCOUR]; ACTRN12614000341628).

Keywords: cancer therapeutics–related cardiac dysfunction; cardiotoxicity; global longitudinal strain; single-view longitudinal strain.