. 2018 Nov 1;27(5):664-670.
doi: 10.1093/icvts/ivy162.

The flutter-by effect: a comprehensive study of the fluttering cusps of the Perceval heart valve prosthesis

Affiliations

The flutter-by effect: a comprehensive study of the fluttering cusps of the Perceval heart valve prosthesis

Ali Aljalloud et al. Interact Cardiovasc Thorac Surg. .

Abstract

Objectives: Sutureless aortic valve prostheses are gaining popularity due to the substantial reduction in cross-clamp time. In this study, we report our observations on the cusp-fluttering phenomenon of the Perceval bioprosthesis (LivaNova, London, UK) using a combination of technical and medical perspectives.

Methods: Between August 2014 and December 2016, a total of 108 patients (69% women) with a mean age of 78 years had aortic valve replacement using the Perceval bioprosthesis (34 combined procedures). All patients underwent transoesophageal echocardiography (TOE) intraoperatively. TOE was performed postoperatively to detect paravalvular leakage and to measure gradients, acceleration time, Doppler velocity indices (Vmax and LVOT/Vmax AV) and effective orifice area indices. In addition, a TOE examination was performed in 21 patients postoperatively. Data were collected retrospectively from our hospital database.

Results: The retrospective evaluation of the intraoperative TOE examinations revealed consistent fluttering in all patients with the Perceval bioprosthesis. The echocardiographic postoperative measurements showed a mean effective orifice area index of 0.91 ± 0.12 cm2/m2. The overall mean pressure and peak pressure gradients were in a higher range (13.5 ± 5.1 mmHg and 25.5 ± 8.6 mmHg, respectively), whereas acceleration time (62.8 ± 16.4 ms) and Doppler velocity indices (0.43 ± 0.11) were within the normal range according to the American Society of Echocardiography or european association of echocardiography (EAE) guidelines. The 2-dimensional TOE in Motion Mode (M-Mode) that was performed in patients with elevated lactate dehydrogenase (LDH) levels revealed remarkable fluttering of the cusps of the Perceval bioprosthesis.

Conclusions: In our study cohort, we observed the fluttering phenomenon in all patients who received the Perceval bioprosthesis, which was correlated with elevated LDH levels and higher pressure gradients.