Hybrid Melody Valve Implantation in Mitral Position in a Child: Usefulness of a 3-Dimensional Printed Model for Preprocedural Planning
- PMID: 29801748
- DOI: 10.1016/j.cjca.2018.02.011
Hybrid Melody Valve Implantation in Mitral Position in a Child: Usefulness of a 3-Dimensional Printed Model for Preprocedural Planning
Abstract
We present the case of a 4-month-old child with atrioventricular canal associated with severe left atrioventricular valve dysfunction who previously underwent 3 surgical valve reconstructions without significant improvement. A Hybrid Melody valve (Medtronic, Minneapolis, MN) insertion was planned. Because of the low weight, the risk of left ventricular outflow tract obstruction was significant and therefore evaluated preprocedurally using a cardiac computed tomography-derived 3-dimensional printed model. In vitro tests showed good anchorage of the valve without subaortic obstruction and the procedure was then achieved with an excellent clinical result.
Copyright © 2018 Canadian Cardiovascular Society. Published by Elsevier Inc. All rights reserved.