Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2018 Jul:56:46-51.
doi: 10.1016/j.clinbiomech.2018.05.012. Epub 2018 May 21.

In-vivo stiffness assessment of distal femur fracture locked plating constructs

Affiliations

In-vivo stiffness assessment of distal femur fracture locked plating constructs

Christopher Parks et al. Clin Biomech (Bristol). 2018 Jul.

Abstract

Background: The purpose of this study was to design and validate a novel stiffness-measuring device using locked plating of distal femur fractures as a model.

Methods: All patients underwent a laterally-based approach, with a bridging locked construct after indirect reduction. A custom and calibrated intraoperative stiffness device was applied and the stiffness of the construct was blindly recorded. Fourteen of twenty-seven patients enrolled with distal femur fractures (AO/OTA 33A and 33C) completed the study. Correlations between stiffness and callus formation, working length, working length/plate length ratio, number of distal locking screws, and fracture pattern were explored.

Findings: Callus and modified radiographic union scale in tibias scores as a linear function of stiffness did not correlate (R2 = 0.06 and 0.07, respectively). Construct working length and working length to plate length ratio did not correlate to stiffness (R2 = 0.18 and 0.16 respectively). A combined delayed and nonunion rate was 14%. Lower extremity measure scores were not statistically different when comparing delayed and nonunion with healed fractures.

Interpretation: The lack of correlation may have been due to the mechanical properties of the plate itself and its large contribution to the overall stiffness of the construct. To our knowledge, clinically relevant stiffness has not been described and this study may provide some estimates. This methodology and these preliminary findings may lay the groundwork for further investigations into this prevalent clinical problem. Other parameters not investigated may play a key role such as body mass index and bone mineral density.

Level of evidence: Diagnostic/Prognostic Level II.

Keywords: Distal femoral fractures; Locking plate; Stiffness.

PubMed Disclaimer

Figures

Figure 1.
Figure 1.
Photo of Custom Device Test Set Up
Figure 2.
Figure 2.
Photo of Instron Tension Test Set-Up
Figure 3.
Figure 3.
Photo of Instron Compression Test Set-Up
Figure 4.
Figure 4.
(A) Force-extension plots for a representative specimen measured using the custom device and using the Instron. The plots are nearly identical. (B). Stiffness determined using the Instron was strongly correlated with stiffness determined using the custom device.
Figure 5.
Figure 5.
Novel stiffness measure device applied to distal femoral locking plate
Figure 6.
Figure 6.
WL and WL/PL versus Stiffness with Linear Regression for AO/OTA 33C fracture pattern.

References

    1. Fouz A, Yeadon AJ, Uhthoff HK. Improved fracture healing with less rigid plates. A biomechanical study in dogs. Clin Orthop Relat Res 1997;339:232–45. - PubMed
    1. Epari DR, Kassi JP, Schell H, et al. Timely fracture-healing requires optimization of axial fixation stability. J Bone Joint Surg AM 2007;89:1575–85. - PubMed
    1. Ganesh VK, Ramakrishna K, Ghista DN. Biomechanics of bone-fracture fixation by stiffness-graded plates in comparison with stainless-steel plates. Biomed Eng Online 2005;4:46. - PMC - PubMed
    1. Button G, Wolinsky P, Hak D. Failure of less invasive stabilization system plates in the distal femur: a report of four cases. J Orthop Trauma 2004;18:565–70. - PubMed
    1. Gardner MJ, Griffith MH, Demetrakopoulos D, et al. Hybrid locked plating of osteoporotic fractures of the humerus. J Bone Joint Surg AM 2006;88:1962–7. - PubMed

Publication types