. 2018 Sep:484:171-178.
doi: 10.1016/j.cca.2018.05.047. Epub 2018 May 24.

Evaluation of the new Sysmex UF-5000 fluorescence flow cytometry analyser for ruling out bacterial urinary tract infection and for prediction of Gram negative bacteria in urine cultures

Affiliations
Free article

Evaluation of the new Sysmex UF-5000 fluorescence flow cytometry analyser for ruling out bacterial urinary tract infection and for prediction of Gram negative bacteria in urine cultures

Rita De Rosa et al. Clin Chim Acta. 2018 Sep.
Free article

Abstract

Background: We evaluated the new flow cytometer UF-5000 with a blue semiconductant laser as a screening tool for ruling out urine samples negative for UTI and its ability to predict Gram negatives in culture.

Methods: Flow cytometry and microbiological analysis were performed on 2719 urine samples, sent to our microbiology laboratory with a request for urine culture.

Results: UF-5000 showed a very good performance in the screening process. Carryover and cross-contamination was negligible. 797 samples were culture positive at a cut-off of ≥105CFU/mL. ROC curve analysis for BACT count demonstrated AUC between 0.973, on 2714 samples, 0.959, on 1516 female samples, and 0.988 on 1198 male samples, respectively. At the cut-off of BACT ≥58/μL AND/OR YLC ≥150/μL, SE was 99.4%, SP 78.2%, PPV 65.4% and NPV 99.7%; false negatives were 0.6%, avoiding unnecessary cultures in 55.5% of specimens. "Gram Neg?" flag predicted Gram negatives in culture with a SE of 81.6% and SP of 93.3%.

Conclusion: The new Sysmex UF-5000 showed high diagnostic accuracy in UTI-screening with a very low rate of false negatives. The instrument is capable of predicting Gram negatives with a good SE and a high agreement with the culture, even if this performance needs further evaluation.

Keywords: Bacteriuria screening; Flow cytometer; UF-5000; Urinalysis; Urinary tract infection (UTI); Urine culture.