Diagnostic assessment of a deep learning system for detecting atrial fibrillation in pulse waveforms

Affiliations

Diagnostic assessment of a deep learning system for detecting atrial fibrillation in pulse waveforms

Ming-Zher Poh et al. Heart. 2018 Dec.

Abstract

Objective: To evaluate the diagnostic performance of a deep learning system for automated detection of atrial fibrillation (AF) in photoplethysmographic (PPG) pulse waveforms.

Methods: We trained a deep convolutional neural network (DCNN) to detect AF in 17 s PPG waveforms using a training data set of 149 048 PPG waveforms constructed from several publicly available PPG databases. The DCNN was validated using an independent test data set of 3039 smartphone-acquired PPG waveforms from adults at high risk of AF at a general outpatient clinic against ECG tracings reviewed by two cardiologists. Six established AF detectors based on handcrafted features were evaluated on the same test data set for performance comparison.

Results: In the validation data set (3039 PPG waveforms) consisting of three sequential PPG waveforms from 1013 participants (mean (SD) age, 68.4 (12.2) years; 46.8% men), the prevalence of AF was 2.8%. The area under the receiver operating characteristic curve (AUC) of the DCNN for AF detection was 0.997 (95% CI 0.996 to 0.999) and was significantly higher than all the other AF detectors (AUC range: 0.924-0.985). The sensitivity of the DCNN was 95.2% (95% CI 88.3% to 98.7%), specificity was 99.0% (95% CI 98.6% to 99.3%), positive predictive value (PPV) was 72.7% (95% CI 65.1% to 79.3%) and negative predictive value (NPV) was 99.9% (95% CI 99.7% to 100%) using a single 17 s PPG waveform. Using the three sequential PPG waveforms in combination (<1 min in total), the sensitivity was 100.0% (95% CI 87.7% to 100%), specificity was 99.6% (95% CI 99.0% to 99.9%), PPV was 87.5% (95% CI 72.5% to 94.9%) and NPV was 100% (95% CI 99.4% to 100%).

Conclusions: In this evaluation of PPG waveforms from adults screened for AF in a real-world primary care setting, the DCNN had high sensitivity, specificity, PPV and NPV for detecting AF, outperforming other state-of-the-art methods based on handcrafted features.

Keywords: atrial fibrillation; ehealth/telemedicine/mobile health; premature ventricular beats.