. 2018 Nov;154(5):1024-1034.
doi: 10.1016/j.chest.2018.04.045. Epub 2018 May 31.

Effectiveness of Reprocessing for Flexible Bronchoscopes and Endobronchial Ultrasound Bronchoscopes

Affiliations

Effectiveness of Reprocessing for Flexible Bronchoscopes and Endobronchial Ultrasound Bronchoscopes

Cori L Ofstead et al. Chest. 2018 Nov.

Abstract

Background: Infections have been linked to inadequately reprocessed flexible bronchoscopes, and recent investigations determined that pathogen transmission occurred even when bronchoscope cleaning and disinfection practices aligned with current guidelines. This multisite, prospective study evaluated the effectiveness of real-world bronchoscope reprocessing methods, using a systematic approach.

Methods: This study involved direct observation of reprocessing methods for flexible bronchoscopes, multifaceted evaluations performed after manual cleaning and after high-level disinfection, and assessments of storage conditions. Visual inspections of ports and channels were performed using lighted magnification and borescopes. Contamination was detected using microbial cultures and tests for protein, hemoglobin, and adenosine triphosphate (ATP). Researchers assessed reprocessing practices, and storage cabinet cleanliness was evaluated by visual inspection and ATP tests.

Results: Researchers examined 24 clinically used bronchoscopes. After manual cleaning, 100% of bronchoscopes had residual contamination. Microbial growth was found in 14 fully reprocessed bronchoscopes (58%), including mold, Stenotrophomonas maltophilia, and Escherichia coli/Shigella species. Visible irregularities were observed in 100% of bronchoscopes, including retained fluid; brown, red, or oily residue; scratches; damaged insertion tubes and distal ends; and filamentous debris in channels. Reprocessing practices were substandard at two of three sites.

Conclusions: Damaged and contaminated bronchoscopes were in use at all sites. Inadequate reprocessing practices may have contributed to bioburden found on bronchoscopes. However, even when guidelines were followed, high-level disinfection was not effective. A shift toward the use of sterilized bronchoscopes is recommended. In the meantime, quality management programs and updated reprocessing guidelines are needed.

Keywords: Stenotrophomonas maltophilia; endoscope; infection; prevention; sterilization.