Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty
- PMID: 29896543
- PMCID: PMC5994790
- DOI: 10.1016/j.artd.2017.10.002
Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty
Abstract
Failures unique to posterior cruciate-substituting total knee prostheses rarely include polyethylene post fractures but have been described. We report a case involving a fracture of the distal insert cone of a rotating-platform (RP) polyethylene liner in a primary total knee arthroplasty. This case highlights a 67-year-old male presenting with new-onset knee pain and recurrent effusions with osteolysis 11 years following placement of a posterior-stabilized, RP total knee arthroplasty. At the time of revision surgery, the polyethylene insert cone was found to be fractured just below the junction between cone and the body of the insert. Liner exchange, synovectomy, and osteolytic-defect curettage and cement packing were performed. One year following revision surgery, the patient is without pain and has returned to function without limitations. Clinicians must be aware of this possible failure with RP prostheses in the setting of pain with a stable knee, recurrent aseptic effusions, and osteolysis.
Keywords: insert stem; post fracture; posterior-stabilized; revision total knee arthroplasty.
Figures





Similar articles
-
Simultaneous Post and Cone Fractures of a Polyethylene Insert in a Mobile-Bearing Posterior-Stabilized Total Knee Arthroplasty.Arthroplast Today. 2024 Feb 16;26:101316. doi: 10.1016/j.artd.2024.101316. eCollection 2024 Apr. Arthroplast Today. 2024. PMID: 38389865 Free PMC article.
-
Nontraumatic Tibial Polyethylene Insert Cone Fracture in Rotating-Platform Total Knee Arthroplasty.Arthroplast Today. 2021 Apr 8;8:283-288.e1. doi: 10.1016/j.artd.2021.02.013. eCollection 2021 Apr. Arthroplast Today. 2021. PMID: 34095405 Free PMC article.
-
Cementless, Cruciate-Retaining Primary Total Knee Arthroplasty Using Conventional Instrumentation: Technical Pearls and Intraoperative Considerations.JBJS Essent Surg Tech. 2024 Sep 13;14(3):e23.00036. doi: 10.2106/JBJS.ST.23.00036. eCollection 2024 Jul-Sep. JBJS Essent Surg Tech. 2024. PMID: 39280965 Free PMC article.
-
Comparison of highly cross-linked and conventional polyethylene in posterior cruciate-substituting total knee arthroplasty in the same patients.J Bone Joint Surg Am. 2014 Nov 5;96(21):1807-13. doi: 10.2106/JBJS.M.01605. J Bone Joint Surg Am. 2014. PMID: 25378508
-
The use of highly cross-linked polyethylene in total knee arthroplasty.J Am Acad Orthop Surg. 2011 Mar;19(3):143-51. doi: 10.5435/00124635-201103000-00003. J Am Acad Orthop Surg. 2011. PMID: 21368095 Review.
Cited by
-
Simultaneous Post and Cone Fractures of a Polyethylene Insert in a Mobile-Bearing Posterior-Stabilized Total Knee Arthroplasty.Arthroplast Today. 2024 Feb 16;26:101316. doi: 10.1016/j.artd.2024.101316. eCollection 2024 Apr. Arthroplast Today. 2024. PMID: 38389865 Free PMC article.
-
Nontraumatic Tibial Polyethylene Insert Cone Fracture in Rotating-Platform Total Knee Arthroplasty.Arthroplast Today. 2021 Apr 8;8:283-288.e1. doi: 10.1016/j.artd.2021.02.013. eCollection 2021 Apr. Arthroplast Today. 2021. PMID: 34095405 Free PMC article.
-
Effect of material selection on tibial post stresses in posterior-stabilized knee prosthesis.Bone Joint Res. 2020 Nov;9(11):768-777. doi: 10.1302/2046-3758.911.BJR-2020-0019.R2. Bone Joint Res. 2020. PMID: 33135462 Free PMC article.
References
-
- Swanik C.B., Lephart S.M., Rubash H.E. Proprioception, kinesthesia, and balance after total knee arthroplasty with cruciate-retaining and posterior stabilized prostheses. J Bone Joint Surg. 2004;86-A:328. - PubMed
-
- Misra A.N., Hussain M.R.A., Fiddian N.J., Newton G. The role of the posterior cruciate ligament in total knee replacement. J Bone Joint Surg. 2003;85:389. - PubMed
-
- Kleinbart F.A., Bryk E., Evangelista J., Scott W.N., Vigorita V.J. Histologic comparison of posterior cruciate ligaments from arthritic and age-matched knee specimens. J Arthroplasty. 1996;11:726. - PubMed
-
- Conditt M.A., Noble P.C., Bertolusso R., Woody J., Parsley B.S. The PCL significantly affects the functional outcome of total knee arthroplasty. J Arthroplasty. 2004;19:107. - PubMed
Publication types
LinkOut - more resources
Full Text Sources
Other Literature Sources